1 |
SIZE |
L710*H660*d560mm |
2 |
Weight |
46 Kg |
3 |
Power |
AC220V 50/60Hz |
4 |
Operational Environment |
0-50 C, 5%-95%Rh |
5 |
Control Method |
Software automatic controlling |
6 |
Communication Methods |
RS-485, upgraded to the USB or network interface |
7 |
Built-in Detector |
Superior sensitive TCD Detector Micro sample cell |
8 |
Carrier Gas |
He, Ar, N2, H2 |
9 |
Carrier Gas purity |
>99.995% |
10 |
Reaction Materials |
NH3, CO, CO2, H2,Methanol, ethanol, pyridine |
11 |
Gas Flow Control |
Digital Mass flow controller (DMFC) control, precision1% |
12 |
Sample Volume |
10-500mg |
13 |
Reactor |
U Type, Quartz or SS316L , ID6mm, OD8mm, Pits or variable diameter |
14 |
High Temperature Furnace |
Stainless steel shell tube furnace,Two structure,max temperature 1100C,Max High heating rate100C/min |
15 |
Gas Circuit Switch |
Germany Bukert magnetic valve, Switzerland VICI Plane switch valve |
16 |
Vapor Generator |
Soft package heating, temperature controller control |
17 |
Constant current source |
0-100mA, 0.1%precision, 1mAstep |
18 |
Detector |
Micro TCD Detector, Min discriminable volume1*10-4cm3/g |
19 |
Precision |
Veracity1%, Repeatability1% |
20 |
Continuous Working |
Long cycle of continuous working |
21 |
Anticorrosion |
Under the optimal situation, it can support acid-resistance, perbasic concentration gas |
22 |
Software |
Special analysis workstation software, base on Windows platform |
23 |
Data Export |
Support TXT Text Output, which makes the exchange data with drawing tools like Origin more convenient |