    HS2000 High Resolution Fiber Spectrometer
     
    Introduction:
    HS2000 high-resolution fiber optic spectrometer is a high-performance fiber optic spectrometer for ultraviolet, visible light, and near-infrared spectrum applications; high-speed USB2.0 host computer operation, the combined module can measure 184 - 640nm/184-840nm, and the resolution is better than 0.1nm@ 200nm, compact size, with sampling accessories, can build a variety of modular spectral test systems; common applications include absorbance and reflectance measurement, water quality testing, light source measurement, fluorescence spectroscopy, LIBS spectroscopy applications, spark spectroscopy applications;
    Features:
    1. Modular design, which can be easily combined according to customer needs
    2. Combined modules can realize wide spectrum (184-620nm/184-840nm)
    3. High resolution 0.1nm@200nm
    4. High-throughput design F-number 3.27
    5. Optical fiber SMA905 interface, convenient for various applications
    Technical parameter:
    Optical table: C-T asymmetric optical path
    Wavelength range: 184-620nm/184-840nm, customized
    Optical resolution (FWHM): 0.1-0.3nm (related to grating and slit)
    Slit size (µm): 10, 25, 50, 100um, customized
    Grating: 3600lines/mm, 2400lines/mm, 1200lines/mm combination
    Optical path optimization: Elimination of second-order diffraction spectrum / second-order cylindrical mirror focusing
    Stray light: 0.04 - 0.1% depending on grating type
    Signal-to-noise ratio: >300:1 at saturation
    Dark Noise: 20RMS counts
    Dynamic Range: 1300:1
    Detector type: Hamamatsu 11639-01 2048 pixel CMOS
    Sensitivity: 1300V/(lx.s)
    Optical interface: standard SMA905
    AD Converter: 16-bit, 2MHz
    Integration time: 10us – 1000s
    Communication interface: USB2.0 high-speed interface, 480Mbps
    Data transfer speed: 1.8ms/sample (USB2.0)
    External trigger: external device TTL synchronization trigger acquisition
    External trigger delay time: 0.3us-100ms (software can be set)
    Power supply form: USB power supply
  • Introduction:
    HS2000 high-resolution fiber optic spectrometer is a high-performance fiber optic spectrometer for ultraviolet, visible light, and near-infrared spectrum applications; high-speed USB2.0 host computer operation, the combined module can measure 184 - 640nm/184-840nm, and the resolution is better than 0.1nm@ 200nm, compact size, with sampling accessories, can build a variety of modular spectral test systems; common applications include absorbance and reflectance measurement, water quality testing, light source measurement, fluorescence spectroscopy, LIBS spectroscopy applications, spark spectroscopy applications;
     
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
+86-18620722006
instruments@instrumentstrade.com