Introduction GC - 9760 special transformer oil portable chromatograph with miniaturization, modularization and digital design concept, the whole machine is small in size, light weight, easy to carry to the scene for rapid analysis; Instrument process with the classic three detector, a sample, 8 minutes to complete the dissolved in insulating oil of H2, CO, CO2, CH4, C2H4, C2H6, C2H2, seven kinds of gas composition content of the whole analysis, and through equipped with dedicated workstations for transformer latent fault analysis and judgment, and analyze the data of its traditional laboratory chromatograph data. Features 1. Electronic flow, according to a pneumatic pressure digital display 2. Adopt computer control technology, realize the instrument temperature Settings, display etc 3. Adopt computer control technology, to realize a key switch on 4. Click of a mouse to complete the ignition and current Settings, and so on 5. Digital FID electronic zero technology, enhance the anti-interference ability of the instrument 6. Adopt computer control technology, the control system for the vibration meter 7. The built-in mute air pump, without external air source 8. Thermal conductivity tungsten expired protection function Technical parameters 1.the minimum detection concentration (PPM)
Component name
H2
CO
CO2
CH4
C2H4
C2H6
C2H2
limit of detection
2
2
2
0.1
0.1
0.1
0.1
Detection range
2-10000
2-50000
2-50000
0.1-2000
0.1-2000
0.1-2000
0.1-2000
2.analysis time: <8min 3.stable time: < 30 min 4.operating temperature: room temperature 8 ℃ ~ 399 ℃ on the accuracy of plus or minus 0.1 ℃ 5.appearance size: 380 mm * 450 mm * 180 mm 6.heavy quantity: < 15 kg 7.electricity source,AC 220 v + 10%, 50 hz - 60 hz 8.the working temperature: - 8 ℃ and 40 ℃ 9. : Communication :USB&RS 232 optional
Introduction GC - 9760 special transformer oil portable chromatograph with miniaturization, modularization and digital design concept, the whole machine is small in size, light weight, easy to carry to the scene for rapid analysis; Instrument process with the classic three detector, a sample, 8 minutes to complete the dissolved in insulating oil of H2, CO, CO2, CH4, C2H4, C2H6, C2H2, seven kinds of gas composition content of the whole analysis, and through equipped with dedicated workstations for transformer latent fault analysis and judgment, and analyze the data of its traditional laboratory chromatograph data.