Measurement items | |
Measurement items | Roundness, cylindricity, concentricity, coaxiality (multi-element), straightness, perpendicularity (multi-element), runout (axial, radial), full runout (axial, radial), planeness (multi-element), parallelism (multi-element), Notch removal, corrugation, spectral analysis, wave height analysis, harmonic analysis, diameter, taper (vertical, plane), wall thickness deviation |
Measuring range | |
Maximum measured diameter | φ400mm |
Maximum measured height | 550mm |
Load capacity | 50Kg |
Spindle precision | |
Spindle radial error H = height of table | ±(0.0125+3H/10000)μm |
Spindle axial error X = distance from spindle center | ±(0.0125+4X/10000)μm |
Staging | |
Table diameter | φ220mm |
Justification range | Adjust eccentricity ±2mm; adjust level ±1 ° |
Tactical diameter | φ400mm |
Spin velocity | 0-12 rpm / min |
Z axis (column) | |
Straightness (narrow range) | 0.35μm/100mm |
Straightness (full range) | 0.6μm/400mm |
Grating resolution | 1μm |
Parallelism between the axis of rotation and the guide rail of Z axis | |
Parallelism between the axis of rotation and the guide rail of Z axis | 1μm /400mm(Bus datum) |
Horizontal arm (transverse guideway) | |
Moving precision | 1.5μm/150mm |
Horizontal moving distance | 180mm |
Grating resolution | 1μm |
Sensor | |
Range | 500μm (Radius difference) |
Resolution ratio | 0.005μm |
Needle shape | 2mm gemstone ball probe (more diameters optional) |
Dynamometry | 1-12g |
Data collection | |
Data collection | 14400 points / week (international) |
Amplified factor | |
Amplified factor | 200000 times automatic |
Evaluation method | |
Evaluation method | LSC (least square), MZC (minimum region method), MIC (maximum tangent circle method), MCC (minimum circumscribed circle method) |
Filter value | |
Filter value | 1-500、1-150、1-50、1-15、3-16、3-17、15-500、17-100 upr |
Data analysis and processing | |
Frequency analysis | Analysis of amplitudes of different Frequency components |
Abnormal data elimination | Manually or automatically remove abnormal data, such as burrs, holes and bumps, etc. |
File management | Automatically save measurement data, can be deleted |
Data printing | Single or multiple measurements can be printed in the same page mode |
Installation requirements | |
Power supply | 220-240v AC, 50HZ, separate ground wire;Maximum power consumption is 450W |
Compressed air | Pure compressed air: pressure 0.4mp, consumption 0.1m3 /min (air source supply 0.6mp, 0.25m3 /min and above) |
Environmental requirements | Working temperature: 10 ° to 30 ° (room temperature).Humidity 20-80%RH (no condensation), good ground line, no obvious source of vibration |