Measurement items |
Measurement items |
Roundness, cylindricity, concentricity, coaxiality (multi-element), straightness, perpendicularity (multi-element), runout (axial, radial), full runout (axial, radial), planeness (multi-element), parallelism (multi-element), Notch removal, corrugation, spectral analysis, wave height analysis, harmonic analysis, diameter, taper (vertical, plane), wall thickness deviation |
Measuring range |
Maximum measured diameter |
φ300mm |
Maximum measured height |
420mm |
Load capacity |
25Kg |
Spindle precision |
Spindle radial error H = height of table |
±(0.025+5H/10000)μm |
Spindle axial error X = distance from spindle center |
±(0.02+6X/10000)μm |
staging |
Table diameter |
φ180mm |
Justification range |
Adjust eccentricity ±2mm; adjust level ±1 ° |
Tactical diameter |
φ300mm |
Spin velocity |
0-12 rpm / min |
Z axis (column) |
Straightness (narrow range) |
0.5μm/100mm |
Straightness (full range) |
0.8μm/300mm |
Grating resolution |
No |
Parallelism between the axis of rotation and the guide rail of Z axis |
Parallelism between the axis of rotation and the guide rail of Z axis |
2μm /300mm(Bus datum) |
Horizontal arm (transverse guideway) |
Moving precision |
2μm/150mm |
Horizontal moving distance |
160mm |
Grating resolution |
No |
Sensor |
Range |
500μm (Radius difference) |
resolution ratio |
0.005μm |
Needle shape |
2mm gemstone ball probe (more diameters optional) |
Dynamometry |
1-12g |
Data collection |
Data collection |
7200 points / week (international) |
Amplified factor |
Amplified factor |
200000 times automatic |
Evaluation method |
Evaluation method |
LSC (least square), MZC (minimum region method), MIC (maximum tangent circle method), MCC (minimum circumscribed circle method) |
Filter value |
Filter value |
1-500、1-150、1-50、1-15、3-16、3-17、15-500、17-100 upr |
Data analysis and processing |
frequency analysis |
Analysis of amplitudes of different Frequency components |
Abnormal data elimination |
Manually or automatically remove abnormal data, such as burrs, holes and bumps, etc. |
File management |
Automatically save measurement data, can be deleted |
Data printing |
Single or multiple measurements can be printed in the same page mode |
Installation requirements |
source |
220-240V AC, 50HZ, single ground wire; maximum power consumption 450W |
compressed air |
Pure compressed air: 0.4MPp, consumed 0.1m3 / min (0.25m3 / min supplied by gas source) |
Environmental requirements |
Working temperature: 10 °-30 °(normal temperature); humidity 20-80%RH (no dew), good ground line, no obvious vibration source |