ATP2001 Refrigerated Micro Fiber Spectrometer Introduction
ATP2001 is developed on the basis of ATP2000. It adds refrigeration technology to the sensor, which can make the CCD work in a set temperature environment. It also has high reliability, ultra-high speed, low cost, high cost performance and other characteristics, and can be adapted to online testing, etc. A miniature spectrometer for various environmental uses.
ATP2001 uses 2048 pixel cooling linear CCD, CCD uses semiconductor refrigeration technology, as long as the external environment of the spectrometer is between -25 degrees and 40 degrees, the CCD can work in a set constant temperature environment (as low as 5oC), thus enabling The ATP2001 achieves extremely high reliability, and the measurement results do not change with the ambient temperature. At the same time, the ATP2001's noise-to-noise ratio is also improved, which is double that of the ATP2000. The ATP2001 can be adapted to the 200-1100nm wavelength range. The CCD detector exposure time can be controlled within 1ms, and the customer can precisely control the signal to noise ratio of the spectrometer.
ATP2001 is ideal for UV, visible, and near-infrared spectroscopy applications. Different slits, gratings, mirrors, and filters are available. Depending on your needs, the spectrometer can be configured for different applications. The spectrum range is from 200nm. Up to 1050nm, the spectral resolution can be chosen from 0.5 to 4.0nm, and the company can also provide OEM customers with customized options.
ATP2001 can receive the SMA905 interface fiber input or free space input of the light to be measured, measure according to the set integration time, and pass the measurement result through USB2.0 (high speed) or UART output; Features
1.Refrigerated detectors, up to 5oC;
2.High noise to noise ratio: 1 time higher than ATP2000;
3.Other parameters are similar to ATP2000;
4.Spectral range: 200-1100 nm;
5.Optical path structure: cross C-T optical path;
6.Detector: 2048 pixel CCD;
7.Integration time: 1ms-130s
8.Power supply: DC 5V@<2A;
9.Power interface: 3.5mm locomotive;
10.ADC bit depth: 16 bits;
11.ADC sampling rate: 2MHz;
12.Optical input interface: SM905 fiber interface or free space input;
13.Data output interface: USB2.0 (high speed) or UART;
14. 20-pin expansion interface; Technical Parameters DETECTOR
Type: Refrigerated linear array CCD, down to 5oC
Detection spectral range: 200-1100 nm
Effective pixels: 2048
Cell size: 14μm × 200μm
Full scale range: ~100 ke-
Sensitivity: 130 Photon @ 400 nm; 60 Photon @ 600 nm
Dark noise: 48.6 RMS @ 0 ̊C; 57.8 RMS @ 13 ̊C (10s Integration)
Wavelength range: 200-400nm, 200-850nm, 200-1000 nm and other wavelength ranges are available, different ranges can be customized
Optical resolution: 0.1-4 nm (depending on slit, spectral range)
SQ: >300:1
Dynamic range: 1300:1
Working temperature: 0-40 oC
Operating humidity: < 90% RH OPTICAL CONFIGURATION
Optical design: f/4 cross-symmetric C-T optical path
Focal length: 40 mm for incidence / 60 mm for output
Incident slit width: 5, 10, 25, 50, 100, 150, 200 μm optional, other sizes can be customized
Incident optical interface: SMA905 fiber interface, free space ELECTRICAL PARAMETER
Integration time: 1 ms - 130 second
Data output interface: USB 2.0
ADC bit depth: 16 bit
Power supply: DC4.5 to 5.5 V (type @5V)
Working current: 170mA@Typ.
Storage temperature: -20°C to +70°C
Operating temperature: 0°C to +50°C PHYSICAL PARAMETERS
Size: 102 × 720 × 34 mm3
Weight: 0.2 kg
Introduction
ATP2001 is developed on the basis of ATP2000. It adds refrigeration technology to the sensor, which can make the CCD work in a set temperature environment. It also has high reliability, ultra-high speed, low cost, high cost performance and other characteristics, and can be adapted to online testing, etc. A miniature spectrometer for various environmental uses.
ATP2001 uses 2048 pixel cooling linear CCD, CCD uses semiconductor refrigeration technology, as long as the external environment of the spectrometer is between -25 degrees and 40 degrees, the CCD can work in a set constant temperature environment (as low as 5oC), thus enabling The ATP2001 achieves extremely high reliability, and the measurement results do not change with the ambient temperature. At the same time, the ATP2001's noise-to-noise ratio is also improved, which is double that of the ATP2000. The ATP2001 can be adapted to the 200-1100nm wavelength range. The CCD detector exposure time can be controlled within 1ms, and the customer can precisely control the signal to noise ratio of the spectrometer.
ATP2001 is ideal for UV, visible, and near-infrared spectroscopy applications. Different slits, gratings, mirrors, and filters are available. Depending on your needs, the spectrometer can be configured for different applications. The spectrum range is from 200nm. Up to 1050nm, the spectral resolution can be chosen from 0.5 to 4.0nm, and the company can also provide OEM customers with customized options.
ATP2001 can receive the SMA905 interface fiber input or free space input of the light to be measured, measure according to the set integration time, and pass the measurement result through USB2.0 (high speed) or UART output;