    ZR-1006A respirator particle filter efficiency tester
    Introduction:
    The ZR-1006A (salt based)/B (oil based) respirator particle filter efficiency and airflow resistance tester can be used to test the filter efficiency and resistance of respirators, filter materials and other particles in medical device inspection centers, safety protection inspection centers, drug inspection centers, disease prevention and control centers, textile inspection centers, hospitals, mask R&D and manufacturers.
    characteristic:
    1. The 8-inch color high-definition LCD touch screen is adopted, which makes the content more intuitive and the operation easier;
    2. Equipped with special saline/oily aerosol generator, which can produce aerosols with specific particle size requirements and concentrations;
    3. It is equipped with a variety of special fixtures, which are suitable for testing various specifications and materials;
    4. It can automatically calculate filtering efficiency and sampling resistance;
    5. Equipped with standard membrane and standard resistance plate;
    6. Built in high-precision electronic flowmeter and high-performance sampling module to ensure flow stability;
    7. Built in long-life photometer module, cumulative sampling time, prompt light path cleaning;
    8. Equipped with aerosol electrostatic neutralization device;
    9. External air source is adopted, with low working noise;
    10. Real time/historical detection data can be exported via USB flash disk or printed by thermal printer.
    Technical parameters:
    Test flow: (30 100) L/min
    Power supply: AC (220 ± 22) V, (50 ± 1) Hz
    Air source: clean cold dry gas ≥ 0.65MPa, ≥ 120L/min
    Host size: (630 long × 650 wide × Height 1670) mm
    Mainframe power consumption:<500W
    Host weight: about 92kg
  • Introduction:
    The ZR-1006A (salt based)/B (oil based) respirator particle filter efficiency and airflow resistance tester can be used to test the filter efficiency and resistance of respirators, filter materials and other particles in medical device inspection centers, safety protection inspection centers, drug inspection centers, disease prevention and control centers, textile inspection centers, hospitals, mask R&D and manufacturers.
     
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
+86-18620722006
instruments@instrumentstrade.com