Technical parameters:
Power supply: AC220V plus or minus 22V / 50Hz;
Power: 3000VA;
Dimensions: 2530*1250*2290mm (length * width * height)
Noise: no more than 70dB (A);
Pressure difference: 0 ~ 80Pa (adjustable);
Cleanliness: class A (static);
Kill rate SAL: 10-6
Hydrogen peroxide: 30% (analytically pure)
Sterilization time: 80min
Operation port: 6 (3+3 double-sided operation)
Differential pressure resolution: 0.1pa
The temperature resolution: 0.1 ℃
Humidity resolution: 0.1%RH
Hour leakage rate: 0.5% of Q/V
High efficiency filter: H14; Technical advantages:
The sterilization unit of vaporized hydrogen peroxide can be integrated to achieve the efficiency of 10-6sal.
With a two-compartment design, the test chamber can be kept in a sterile state and work efficiency is doubled.
Double-sided transparent window, double-sided operation, 3+3 glove mouth design, no operation blind area and visual Angle blind area.
Continuous garbage effluent channel, using RTP aseptic transfer technology, does not damage the sterile environment in the cabin.
Integrated peristaltic pump, easy to operate, does not take up space.
Description of basic functions and features
Siemens PLC programmable controller program automatic control;
Siemens touch liquid crystal screen human-machine dialogue interface, humanized design;
Third level administrator password access;
UBS interface data export and print function;
Automatic and manual operation mode;
Temperature, humidity and pressure difference monitoring function;
Pressure loss alarm function (audible and visual signals);
Waterproof and sterile socket in the cabin (company patent);
Door interlock function, special condition limit open function.
Technical advantages:
The sterilization unit of vaporized hydrogen peroxide can be integrated to achieve the efficiency of 10-6sal.
With a two-compartment design, the test chamber can be kept in a sterile state and work efficiency is doubled.
Double-sided transparent window, double-sided operation, 3+3 glove mouth design, no operation blind area and visual Angle blind area.
Continuous garbage effluent channel, using RTP aseptic transfer technology, does not damage the sterile environment in the cabin.
Integrated peristaltic pump, easy to operate, does not take up space.
Description of basic functions and features
Siemens PLC programmable controller program automatic control;
Siemens touch liquid crystal screen human-machine dialogue interface, humanized design;
Third level administrator password access;
UBS interface data export and print function;
Automatic and manual operation mode;
Temperature, humidity and pressure difference monitoring function;
Pressure loss alarm function (audible and visual signals);
Waterproof and sterile socket in the cabin (company patent);
Door interlock function, special condition limit open function.