1
Specification model | Winner2006A | Winner2006B | ||
Executive Standard | ISO13320-1:1999,GB/T19077.1-2008,Q/0100JWN001-2013 | |||
Test Range | 0.01-780μm | 0.1-1000μm | ||
Number of detector channels | 68 | 49 | ||
Accuracy error | <0.5% (D50 value of national standard sample) | |||
Repeatability error | <0.5% (D50 value of national standard sample) | |||
Laser parameters | Main laser source: US imported semiconductor fiber laser λ = 650nm, p> 2mW Auxiliary laser source: semiconductor laser, λ = 650nm, p> 2mW, service life:> 25000H |
|||
Built-in dispersion method | Ultrasound | Frequency: f = 40KHz, power: p = 35W, time: ≥1S | ||
Stir | Speed: 0-3000rpm adjustable speed | |||
cycle | Rated flow: 10L / min Rated power: 15W | |||
Sample cell | Capacity: 450mL | |||
Operating mode | Software operation / fully automatic operation mode can be switched | |||
Automatic alignment system | Automatic alignment system | |||
Software function | Analysis mode | Including free distribution, R-R distribution and logarithmic normal distribution, grading statistical mode according to mesh, etc., to meet the different requirements of different industries on the statistical method of sample size measurement. | ||
Statistical methods | Volume distribution and quantity distribution to meet different statistical methods for particle size distribution in different industries. | |||
Statistical comparison | Statistical comparison and analysis can be performed on multiple test results, which can clearly compare the differences between different batches of samples, samples before and after processing, and test results at different times, which has a strong practical significance for the quality control of industrial raw materials. | |||
Custom analysis | User-defined analysis parameters, find percentage based on particle size, calculate particle size according to percentage, or calculate percentage according to particle size interval, to meet the characterization of particle size testing in different industries. | |||
testing report | The test report can export documents in various forms such as Word, Excel, Picture (Bmp), and Text (Text) to meet the test results cited in the test report and scientific research articles in any occasion. | |||
Multilingual support | Chinese and English language interface support, and other language interfaces can also be embedded according to user requirements. | |||
Intelligent operation mode | With intelligent operation mode, it can automatically control the steps of water intake, dispersion, and testing, which not only reduces the workload of testers, but also has better repeatability of test results due to no human factors. | |||
Test speed | <2min / time | |||
volume | 850mm*390mm*450mm | |||
weight | 40Kg | |||