TC830 coating thickness gauge Introduction
TC830 coating thickness gauge can quickly, damage-free and accurately measure the thickness of coatings and coatings. TC830 coating thickness gauge can be used in both laboratory and engineering field. TC830 coating thickness gauge can be widely used in manufacturing, metal processing industry, chemical industry, commodity inspection and other inspection fields. It is an indispensable instrument for material protection professionals. Features
1. Use magnetic sensor to measure the thickness of non-magnetic coating on magnetic metal substrate
Eddy current sensor can be used to measure the thickness of non-conductive coating on non-magnetic metal substrate
2. With two measurement modes: high precision single measurement mode (Precision) and scan mode (Scan)
3. With temperature compensation function: advanced real-time temperature compensation technology can automatically
Compensate for measurement errors to make measurements more accurate
4. After the F-type magnetic sensor is worn, it can be automatically matched with the main unit of the instrument. After successful matching and basic calibration,
Continued use to extend sensor life
5. Set five statistics: average (MEAN), maximum (MAX), minimum (MIN), number of tests (NO.),
Standard deviation (S.DEV)
6. The instrument can be calibrated by zero-point calibration, one-point calibration or two-point calibration, and basic calibration and temperature system
Number calibration method to correct the system error of the sensor
7. With storage function: can store up to 500 measured values
8. With delete function: delete the single suspicious data in the measurement, and also delete all the data in the storage area,
For new measurements
9. Limits can be set: automatic alarm for measured values outside the limits
10.With power and power display function
11. Buzzer sounds during operation, and you can choose to turn off the buzzer
12. There are three types of shutdown methods: manual shutdown mode, automatic shutdown mode when overtime and automatic shutdown mode with low battery.
Set the timeout for automatic shutdown after overtime, with backlight flashing and beeping sounds during automatic shutdown
13. With real-time data communication function: each measurement result can be sent to a PC for editing or printing in real time Technical Parameters
Model: TC830FN Magnetic / Eddy Current
Working principle: TC830F magnetic induction, TC830N eddy current
Measuring range: 0 ~ 1500um (other specifications can be customized)
Resolution: 0.1um (when the measurement range is less than 100um)
1um (when the measurement range is greater than 100um)
Display point calibration: ± (2% H + 1)
Indication two-point calibration: ± [(1 ~ 2)% H + 1]
Basic calibration: ± (1% H + 1)
Minimum measuring area: Φ7mm, Φ5mm
Thinnest measurement base: 0.5mm, 0.3mm
Minimum curvature radius of substrate: convex: 1.5 concave: 10mm, convex: 3 concave: 10mm
Metric / Imperial: Metric and English units can be selected
Battery voltage indication: Low voltage alert function
Automatic shutdown: with automatic shutdown function
With data communication: each measurement result can be sent to a PC for printing in real time
Power source: Three 1.5V AAA batteries
Operating environment: Temperature: 0 ~ 40 ℃ Humidity: 10 ~ 90% RH
Dimensions: 110 × 50 × 23mm
Weight: 100g (without battery)
Introduction
TC830 coating thickness gauge can quickly, damage-free and accurately measure the thickness of coatings and coatings. TC830 coating thickness gauge can be used in both laboratory and engineering field. TC830 coating thickness gauge can be widely used in manufacturing, metal processing industry, chemical industry, commodity inspection and other inspection fields. It is an indispensable instrument for material protection professionals.