Introduction: RT1903A Glove Integrity Tester is a special instrument used to test the integrity of isolator/RABS system sleeve/glove or integrated glove. The tester includes a control panel that is convenient for the operator, a control system, and a test port that can be matched with the isolator operation port. The tester can work in positive pressure mode. The test result is displayed as the pressure drop value per unit time. The operator determines the integrity of the testing gloves based on the reference values provided by the manufacturer. Application areas: Used to test the integrity of the isolator/RABS system sleeve/glove or one-piece glove Features: 1. Developed according to ISO14644-7 standard; 2. All stainless steel case, easy to clean, suitable for use in clean areas; 3. The inclined control panel is easy to observe and operate; 4. The test gas is external compressed air and nitrogen; 5. Provide result printing function to facilitate record keeping; 6. Use touch screen for control, with friendly human-computer interaction interface; 7. The sleeve-gloves can be leak tested before or after using the isolator; 8. Portable structure design, easy to use. Technical Parameters: RT1903A glove integrity tester Input pressure: 0.2MPa~0.6MPa Inflation pressure: 500~1000Pa Test time: 5~8min Pressure display resolution: 0.1Pa Detection accuracy (minimum aperture): 500μm Instrument weight: 20kg Overall dimensions of the host: 530mm (length) × 235mm (width) × 330mm (height) Test port specifications: Ф255mm (other specifications can be customized) Printing function: needle type micro printer Store data: 10 groups Power supply: 220V, 50Hz Power: 200W
Introduction: RT1903A Glove Integrity Tester is a special instrument used to test the integrity of isolator/RABS system sleeve/glove or integrated glove. The tester includes a control panel that is convenient for the operator, a control system, and a test port that can be matched with the isolator operation port. The tester can work in positive pressure mode. The test result is displayed as the pressure drop value per unit time. The operator determines the integrity of the testing gloves based on the reference values provided by the manufacturer.