- Model NumberRC-600C
- Brand NameTZSRCFXYQ
- Payment TermsT/T, paypal
Brief introduction:
The instrument adopts the principle of microcoulomb titration. When the sample to be tested or the standard sample enters the salt electrolysis cell, the Cl- in the sample reacts with the Ag+ in the electrolyte, and the Cl- and the salt content of the sample can be calculated by calculating the electricity consumption of the electrogenerated Ag+ according to Faraday's law.
Scope of application: it is applicable to the determination of salt content in crude oil and its products, and can be used for the determination of salt content in heavy oil, residual oil, refinery and oilfield, various industrial water and discharge water, as well as the measurement of inorganic chloride ion in the above samples.
Conform to the standards: SY/T0536 and other standards.
Equipment integrity: host (amplifier, stirrer), centrifuge, mixer, water bath, computer (including operating software), printer, etc.
Features:
Windows operating platform, with man-machine dialogue function, easy to operate.
The screen displays parameters, measurement peak and measurement results, and can be stored or printed.
Special ceramic sand core is adopted in the electrolytic cell to replace the former Yin and Yang film and improve the measurement sensitivity.
The results were more stable, accurate and reproducible.
Sampling stainless steel motor cap, never oxidize
Technical parameters:
Measurement range: salt content: 0.1 ~ 10000mgNaCl/L (high content can be diluted)
Inorganic chloride ion: 0.1 PPM ~ 100% content (high content can be diluted)
Measurement accuracy: 2mg/L absolute deviation: 0.2mg/L
> 2mg/L relative deviation: 5%
Sample size: salt content: 1g 0.2g
Chloride ion: 1-- 500ul