LPC-L2 oil particle counter (desktop) Introduction
The LPC-L2 oil particle counter is an upgraded product based on LPC-L, which is based on standards such as GB/T 18854-2002 (ISO11171-1999). Developed by the principle of light resistance (shading) counting, it fully complies with the corresponding national military standards, national standards and international standards. It can provide fast, accurate, reliable and repeatable test results and complete pollution monitoring and analysis report. It is suitable for the detection of particle contamination such as hydraulic oil, lubricating oil, anti-fuel, insulating oil and turbine oil. Can be widely used in aviation, aerospace, electric power, petroleum, chemical, transportation, port, metallurgy, machinery, automobile manufacturing and other fields. Features
Using the principle of photoresist (shading) method developed by the International Hydraulic Standards Committee
High-precision laser sensor with wide test range, stable performance, low noise and high resolution
Using high-pressure syringe pump sampling method, you can set the sampling volume by yourself, the injection speed is stable, and the sampling accuracy is high.
Positive and negative pressure combined injection system for sample degassing, suitable for different viscosity test
Built-in pressure sensor, can set the pressure value, and automatically determine the air pressure inside the cabin to ensure safety
Built-in air purification system to ensure testing is not polluted
Built-in multiple calibration curves for calibration with all common standards at home and abroad
Built-in GJB-420B, NAS1638, ISO4406 (GB/T14039), SAE4059E and ГOCT17216-71 and other common standards, support custom standard testing, and set the required standards according to customer needs
Built-in data analysis system, one test can give all standard test data and pollution level
The particle size can be arbitrarily set, and nearly 10,000 particles are built in for easy particle size analysis.
A variety of sampling containers, such as standard sampling bottles or sampling cups, can be used to meet the testing requirements of different industries.
Full-featured automatic color touch screen operation, support Chinese and English dual interface, easy to operate
Single channel and multi-channel calibration for automatic calibration
RS232 interface for connection to a computer or lab platform for data processing
With massive data storage and printing functions, it can store 1000 sets of data and support U disk to store data. Technical Parameters
Light source: semiconductor laser
Particle size range: 0.8um~600um (depending on the sensor)
Detection channel: 16 channels, arbitrarily set particle size
Resolution: better than 10%
Repeatability: RSD<2%
Viscosity range: up to 350cSt
Sampling volume: 0.2~1000ml
Sampling accuracy: better than ±0.5%
Sampling speed: 5 ~ 80mL / min
Maximum pressure in the pressure chamber: 0.08MPa
Maximum positive pressure of the pressure chamber: 0.8 MPa
Maximum particle concentration: 20000 particles / mL
Power supply: 110~245V AC, 50Hz, 70W
Dimensions: 340mm ╳ 410mm ╳ 650mm
Introduction
The LPC-L2 oil particle counter is an upgraded product based on LPC-L, which is based on standards such as GB/T 18854-2002 (ISO11171-1999). Developed by the principle of light resistance (shading) counting, it fully complies with the corresponding national military standards, national standards and international standards. It can provide fast, accurate, reliable and repeatable test results and complete pollution monitoring and analysis report. It is suitable for the detection of particle contamination such as hydraulic oil, lubricating oil, anti-fuel, insulating oil and turbine oil. Can be widely used in aviation, aerospace, electric power, petroleum, chemical, transportation, port, metallurgy, machinery, automobile manufacturing and other fields.