1 Sales LC-100 Liquid Chromatograph,Manufacturer
Position: Home > > Analytical instruments > Chromatogram > Liquid chromatography > LC-100 Liquid Chromatograph for mixture separation test

LC-100 Liquid Chromatograph for mixture separation test

Product ID:FSye008

Price: Please inquire PDF Format
Supply Ability:20SETS/MON
Port:SHENZHEN
share:
  • Product Introduction
  • Consulting
  • Model NumberLC-100
    Brand NameYIMA
    Payment TermsT/T, paypal
    Introduction
    1. LC-P100 high pressure constant flow pump
    (1) Design of reciprocating pump in parallel, The characteristics of high precision flow, small pressure pulsation.Has the flow of high precision, the characteristics of pressure fluctuation is small, in the same infusion quantity under the premise of the service life of sealing ring twice the length of a series pump.
    (2) Original one-way valve design, by the valve core body, gem ball seat and gem ball form a whole structure, it has a simple structure, good sealing, the pump flow rate precision is improved and the test precision of the system.
    (3) Plunger rod automatic indentation function, Don't open the case it's easier to realize seal replacement.
    (4) Adopts imported line and connector, excellent compatibility, Can the greatest degree avoid abnormal work due to the piping system of the pump.
    (5) Using the buffer salt system users, the product provides a liquid pump, double sealing ring design.Delay sealing ring wear, prolong service life.
    (6) Semipreparative pump adn four yuan low pressure system for the user to choose from.
    2.LC - UV100 ultraviolet detector
    (1) The full digital switching system, to avoid the general uv detector of chromatographic signal need multiple mode - a change of signal distortion and interference.
    (2) Flow cell USES the design of parallel double cone hole, improve the SNR, detection effect is better.
    (3) 2000 hours Import deuterium lamp , longer service life and provide more detection sensitivity.
    (4) Optional tungsten light source, universal coverage scope of visible light detection
    3. LC - WS100 chromatography workstation 
    (1) Gradient elute: User input in the corresponding list various ratio of mobile phase, flow rate and the change of time, can be easy and accurate implementation of the operation of gradient elution. Software provides the corresponding pump flow curve, in order to facilitate users to proofread gradient Settings, and the user can more intuitive understanding of the gradient operation process.
    (2) Spectral scanning: Users can directly detect different wavelengths of deuterium the size of the light is strong, and can judge whether the deuterium lamp needs to be replaced.486.6 nm and 656 nm on the map the location of the two to determine the accuracy of the wavelength error value.
    (3) Wavelength scanning: Provided by the wavelength scanning capabilities ranging between 180-680 nm, the minimum step 1 nm, it can help users understand to the samples with different wavelength ultraviolet absorbance, users' detection of different samples with different wavelength to achieve the best effect of analysis.
    (4) Relay events: Can realize the function of various user needs, such as switching to equal flow valve to realize the separation recycling. 
    4. Column incubator (options) 
    Temperature control parts and solvent tray integrated design, function more practical, external appearance more beautiful.2 root chromatographic column can be placed at the same time, can satisfy the series analysis of chromatographic column.
    Technical parameters:
    1. LC-P100 High pressure constant flow pump technical indicator:
    flow range: 0.001~9.999ml/min To 0.001 ml/min step flow adjustment
    Flow stability error: less than 0.3%
    Flow value error: less than ±2%
    Maximum working pressure: 42MPa(0.001~9.999ml/min)
    Pressure is linear and accuracy: Pressure error less than ±0.5 MPa(0~42MPa)
    pressure fluctuation: less than 0.1MPa(flow 1ml/min, pressure 5~10MPa)
    sealing of the pump : pressure 42MPa time 10min, Pressure drop is less than 0.5MPa
    external dimension: 450 mm * 300 mm * 160 mm (length x width x height)
    2.LC-UV100 ultraviolet detector technical parameters:
    Wavelength range: 190 ~ 680 nm
    The baseline noise: ≤2×10-5AU peak - peak (methanol, 1 ml/min, 254 nm, 20 )
    Baseline noise: ≤±0.25×10-5AU (empty pool, response time 1s, 20 )
    Baseline drift: ≤2×10-4AU/hr (methanol, 1 ml/min, 254 nm, 20 )
    Baseline drift: ≤0.4×10-4AU/hr (empty pool, response time 1s, 20 )
    Minimum quantity: ≤5×10-9g/ml (Napthaline/methanol solution)
    Wavelength error value : ≤±2nm
    wavelength repeatability: better than 0.1 nm
    spectral bandwidth: 8nm
    Linear range: better than 104

    Flow cell volume: 8µL
    Optical path: 8 mm
    Shape size: 450 mm * 300 mm * 160 mm (length x width x height) 
    3. LC—CO100 Column incubator technical parameters:
    Temperature control range: room temperature 5  ~ 80 
    Temperature setting range: 10 ~ 80  (1  step)
    Temperature control precision: ±0.1
     

     
  • Introduction
    1. LC-P100 high pressure constant flow pump
    (1) Design of reciprocating pump in parallel, The characteristics of high precision flow, small pressure pulsation.Has the flow of high precision, the characteristics of pressure fluctuation is small, in the same infusion quantity under the premise of the service life of sealing ring twice the length of a series pump.
    (2) Original one-way valve design, by the valve core body, gem ball seat and gem ball form a whole structure, it has a simple structure, good sealing, the pump flow rate precision is improved and the test precision of the system.
    (3) Plunger rod automatic indentation function, Don't open the case it's easier to realize seal replacement.
    (4) Adopts imported line and connector, excellent compatibility, Can the greatest degree avoid abnormal work due to the piping system of the pump.
    (5) Using the buffer salt system users, the product provides a liquid pump, double sealing ring design.Delay sealing ring wear, prolong service life.
    (6) Semipreparative pump adn four yuan low pressure system for the user to choose from.
    2.LC - UV100 ultraviolet detector
    (1) The full digital switching system, to avoid the general uv detector of chromatographic signal need multiple mode - a change of signal distortion and interference.
    (2) Flow cell USES the design of parallel double cone hole, improve the SNR, detection effect is better.
    (3) 2000 hours Import deuterium lamp , longer service life and provide more detection sensitivity.
    (4) Optional tungsten light source, universal coverage scope of visible light detection
    3. LC - WS100 chromatography workstation 
    (1) Gradient elute: User input in the corresponding list various ratio of mobile phase, flow rate and the change of time, can be easy and accurate implementation of the operation of gradient elution. Software provides the corresponding pump flow curve, in order to facilitate users to proofread gradient Settings, and the user can more intuitive understanding of the gradient operation process.
    (2) Spectral scanning: Users can directly detect different wavelengths of deuterium the size of the light is strong, and can judge whether the deuterium lamp needs to be replaced.486.6 nm and 656 nm on the map the location of the two to determine the accuracy of the wavelength error value.
    (3) Wavelength scanning: Provided by the wavelength scanning capabilities ranging between 180-680 nm, the minimum step 1 nm, it can help users understand to the samples with different wavelength ultraviolet absorbance, users' detection of different samples with different wavelength to achieve the best effect of analysis.
    (4) Relay events: Can realize the function of various user needs, such as switching to equal flow valve to realize the separation recycling. 
    4. Column incubator (options) 
    Temperature control parts and solvent tray integrated design, function more practical, external appearance more beautiful.2 root chromatographic column can be placed at the same time, can satisfy the series analysis of chromatographic column.
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
+86-18620722006
instruments@instrumentstrade.com