Introduction: JL-500 extensometer calibration instrument is my company reference to similar products at home and abroad on the basis of today's cutting-edge technology with the development of measuring instruments. It consists of rack, import module, high-precision digital micrometer and other components. Accordance with the JJG762-92 extensometer test procedures specified, dedicated to all kinds of extensometer calibration. Features: 1, with the import module instead of ordinary slider, remove the movement error; 2, high-precision digital dial indicator instead of engraved, intuitive and convenient, eliminating human error. Technical Parameters: 1, measuring extensometer gauge range: 20-500mm, 2, measuring range: 12.7mm or 25.4mm 3, display resolution: 0.001mm, 4, the overall accuracy of ± 3um. 5, reference standard: JJF1096-2002 Extensometer calibration standard ISO 9513: 1999 Metallic materials - Single axis test with extensometer calibration JJG762-1992 Extensometer test procedures 1 extensometer calibration Main configuration: 1, rack, module, dial gauge, a set of poles 2, character manual 1 copy 3, 1 aluminum box
Introduction: JL-500 extensometer calibration instrument is my company reference to similar products at home and abroad on the basis of today's cutting-edge technology with the development of measuring instruments. It consists of rack, import module, high-precision digital micrometer and other components. Accordance with the JJG762-92 extensometer test procedures specified, dedicated to all kinds of extensometer calibration.