ICP-6810 full spectrum direct reading inductively coupled plasma optical emission spectrometer Introduction
ICP-6810 is a full-spectrum direct-reading inductively coupled plasma optical emission spectrometer for the determination of trace and trace elements in different substances (soluble in nitric acid, hydrochloric acid, hydrofluoric acid, etc.). It is widely used in environmental protection, petroleum products, Rare earth, semiconductor, geology, metallurgy, chemical, clinical medicine, food, biological samples, criminal science, agricultural research and other fields. Features
Stable performance
It adopts all solid-state RF power supply, which has many advantages such as small size, high efficiency, stable output power, various protection functions, etc. The load adopts automatic matching technology, and the matching speed is fast, which improves the use efficiency of the power supply and the stability of the instrument. And make the whole ignition process simple and convenient.
Injection automation
With four-channel automatic design, the speed can be adjusted according to the test requirements. The carrier gas, plasma gas and auxiliary gas are controlled by advanced mass flow controller. At the same time, it can be equipped with imported high-salt atomizer and imported hydrogen fluoride-resistant mist. Chemicals, etc., to meet the various testing needs of customers.
Accurate analysis
Echelle grating-prism cross-dispersion, no moving optics; ultra-low stray light design with unique optical design, nitrogen distributed blowing, imported optics, intelligent and accurate automatic wavelength calibration algorithm.
Quick test
American thermoelectric CID detectors, continuous coverage in the 165-900nm range, read the intensity integral value of all analytical lines in one exposure to speed up the analysis. Technical Parameters
Output power: 800 ~ 1500W
Total argon consumption: ≤ 14L/min
Wavelength range: 165nm~900nm
Resolution: <0.0068nm@200nm
Instrument technical indicators
Observation method: vertical observation
Liquid content: 0.01ppm to several thousand ppm
Solid content: 0.001%~70%
Repeatability: RSD<1%
Stability: RSD <1.5% (within 2 hours)
Element detection limit (μg/L): most elements 1ppb~10ppb