1 Sales GCOM-3000 volatile organic compound online analyzer,Manufacturer
Position: Home > > Environmental monitoring and analysis > Gaseous pollutants > VOC/TVOC detector > GCOM-3000 volatile organic compound online analyzer

GCOM-3000 volatile organic compound online analyzer

Product ID:HQWRvoc035

Price: Please inquire PDF Format
Supply Ability:50 SETS/MON
Port:SHENZHEN
share:
  • Product Introduction
  • Consulting
  • Model NumberGCOM-3000
    Brand NameZETIANCHUNLAI
    Payment TermsTT,PAYPAL
    GCOM-3000 volatile organic compound online analyzer
    Introduction:
    The GCOM-3000 volatile organic compound online analyzer uses a dedicated chromatographic column combination, gas path backflushing technology, and hydrogen flame ionization detector (FID) technology to achieve a FID detector for benzene series compounds, suitable for high temperature, high humidity, and high dust conditions, and the data can be uploaded to the corresponding monitoring platform according to requirements. Application field: Online continuous monitoring of VOC in urban atmospheric environment air, pharmaceutical and chemical industries, textile printing and dyeing, furniture manufacturing, automotive spraying and other industrial park boundaries.
    Features:
    1. Use Electronic Mass Flow Control (MFC) to improve the measurement accuracy and stability of the analyzer;
    2. The use of silanized gas pipeline reduces the adsorption residue of gas samples inside the pipe wall;
    3. The intelligent FID detector has a self ignition alarm function to effectively control the safety of the hydrogen flame gas path;
    4. The analyzer supports TCP/IP network for remote control of the analyzer, and can also be operated directly through the touch LCD screen.
    5. Built in sampling pump, active sampling, can directly analyze atmospheric or negative pressure samples without the need for an external booster pump.
    Technical parameters:
    PrincipleGC-FID
    Temp stability of column box≤0.5%
    Measuring rangeNMHC: 0~1000μmol/mol (customizable)
    Measurement periodNMHC≤2mi BTEX≤6min
    Baseline noise≤0.5pA
    Dimension(W×H×D)482×221×559 (mm) (W×H×D)
    Baseline drift≤5pA (30min)
    DisplayTouch screen
    Qualitative repeatabilityRSD7≤1.0%
    Communication output10/100M ethernet port, RS232
    Quantitative repeatabilityRSD7≤2.0%
    Communication protocolModbus-RTU, Hj212 or local protocol
    Span drift:≤3.0% (4h)
    Power220±22VAC, 50±1Hz, 900W
    LODNMHC (measure in carbon)≤0.05ppm
    LinearityR≥0.99
    LOD≤0.1ppm (measure in benzene)
    Temp accuracy of column box±0.1
     

     
  • Introduction:
    The GCOM-3000 volatile organic compound online analyzer uses a dedicated chromatographic column combination, gas path backflushing technology, and hydrogen flame ionization detector (FID) technology to achieve a FID detector for benzene series compounds, suitable for high temperature, high humidity, and high dust conditions, and the data can be uploaded to the corresponding monitoring platform according to requirements. Application field: Online continuous monitoring of VOC in urban atmospheric environment air, pharmaceutical and chemical industries, textile printing and dyeing, furniture manufacturing, automotive spraying and other industrial park boundaries.
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
+86-18620722006
instruments@instrumentstrade.com