DR-2020 multi-parameter heavy metal online analyzer Introduction
Using the principle of anodic stripping voltammetry (ASV) approved by international authoritative institutions, the analytical solution is first pre-electrolyzed under certain conditions to concentrate the component to be measured on the working electrode, and then the solution is allowed to stand still for a period of time, and then the anti-application is applied. The test object enriched in the working electrode is eluted with a voltage, and the concentration of the heavy metal to be tested is determined by the elution voltammetry curve. Features
With a 10.4 inch large touch screen, you can use the instrument without professional training;
The instrument has a small reaction system and supports the separation of waste water and waste liquid.
Small size for easy transportation and installation;
Customized solutions can be provided to reduce user operation and maintenance costs according to water conditions;
Daily maintenance can be completed in 5 minutes;
WeChat applet "Water Quality Cloud" supports remote diagnosis and remote operation and maintenance, WeChat scans two-dimensional code report faults, and the after-sales service is more accurate and faster;
Meet the new technical requirements of national surface water automatic monitoring instruments Technical parameters
Measurement parameters: Zn, Cd, Pb, Cu
Measuring principle: anodic stripping voltammetry (ASV)
Measuring range: 0~0.04mg/L, 0~0.2mg/L, 0~1mg/L
(can be expanded as needed)
Accuracy: ±10% (0.02 ug/L or more); ±2 ug/L (0.02 ug/L or less)
Repeatability: ±5%
Zero drift: ±5%
Range drift: ±5%
Detection limit: Cd≤0.5μg/L; Pb≤1μg/L; Cu≤0.5μg/L; Zn≤1μg/L
Actual water sample comparison test: ±15% (0.02 ug/L or more); ±2ug/L (0.02 ug/L or less)
Spike recovery: ±15%
MTBF: ≥720h/time
Measurement time: ≤45min
Reagent replacement cycle: two weeks
Interface: 4-20mA, RS232, RS485, MODBUS communication protocol
Relay
Using the principle of anodic stripping voltammetry (ASV) approved by international authoritative institutions, the analytical solution is first pre-electrolyzed under certain conditions to concentrate the component to be measured on the working electrode, and then the solution is allowed to stand still for a period of time, and then the anti-application is applied. The test object enriched in the working electrode is eluted with a voltage, and the concentration of the heavy metal to be tested is determined by the elution voltammetry curve.