1 Sales DM1240 X-ray fluorescence calcium iron sulfur analyzer,Manufacturer
Position: Home > > Analytical instruments > Element analyzer > Calcium iron coal analyzer > DM1240 X-ray fluorescence calcium iron sulfur analyzer for Cement Analysis

DM1240 X-ray fluorescence calcium iron sulfur analyzer for Cement Analysis

Product ID:FYgtm004

Price: Please inquire PDF Format
Supply Ability:20SETS/MON
Port:SHENZHEN
share:
  • Product Introduction
  • Consulting
  • Model NumberDM1240
    Brand NameYIMA
    Payment TermsT/T, paypal
    Introduction
     DM 1240 X-ray fluorescence analyzer calcium and iron sulfur is mainly used for cement raw material , clinker and cement and other materials in SO3, CaO, Fe2O3 content analysis in the company's original DM Series calcium and iron analyzer , X fluorescence measurement of sulfur developed on the basis of instrument, such as the launch of the country's first integrated calcium and iron sulfur analyzer. Analytical speed, high precision , human error , low labor intensity operatives , no pollution. DM 1240 X-ray fluorescence analyzer sulfur calcium and iron, in addition to maintaining the original DM series of X- ray fluorescence analyzer has the advantages, made the following significant improvements , making the company X-ray fluorescence analyzer to maintain a leading position .
    Features
    1. No radioactive isotope , non-polluting.
    2. Physical analysis methods, does not destroy the sample contact , green
    3. The three components of the rapid analysis of 30 seconds to give the fastest results .
    4. can be used independently , but also through a standard RS232 or RS485 serial port and batching system for online use .
    5. High degree of integration, environmental adaptability , strong anti-interference ability, high reliability.
    6. Large -screen color touch screen.
    7. audible alarm, interactive manner.
    8. The software features a strong , qualified rate statistics , the rate of value calculation , error message , etc., can also add software features based on user requirements .
    Technical parameters:
    1. The scope of analysis : SO3, CaO, Fe2O3 analysis range can be adjusted by the calibration curve of the selected method .
    2. Analysis Range Width : SO3 max-SO3 min≤5%, CaOmax-CaOmin≤7%, Fe2O3 max-Fe2O3 min ≤5%.
    3. Linearity error : SO3: ≤0.11%, CaO: ≤0.18%, Fe2O3: ≤0.10%.
    4. System Analysis Time : 30 × n seconds (n is a natural number ) , the recommended value is 60 seconds .
    5. Analysis of Accuracy : S SO3≤0.05%, SCaO≤0.09%, S Fe2O3≤0.05%.
    6. Temperature stability : in the range of 5 ~ + 40 ℃, absolute drift : |  SO3|≤0.05%, | CaO |≤0.09%, |  Fe2O3% |
    7. Conditions of Use: Ambient temperature : 5 ~ + 40 ℃, relative humidity : ≤85% (30 ℃), Power supply : 220V ± 20V, 50Hz.
    8. Total power consumption : ≤50W.
    9. Size and weight : 470mm × 365mm × 130mm, 15kg.

     
     
  • Introduction
     DM 1240 X-ray fluorescence analyzer calcium and iron sulfur is mainly used for cement raw material , clinker and cement and other materials in SO3, CaO, Fe2O3 content analysis in the company's original DM Series calcium and iron analyzer , X fluorescence measurement of sulfur developed on the basis of instrument, such as the launch of the country's first integrated calcium and iron sulfur analyzer. Analytical speed, high precision , human error , low labor intensity operatives , no pollution. DM 1240 X-ray fluorescence analyzer sulfur calcium and iron, in addition to maintaining the original DM series of X- ray fluorescence analyzer has the advantages, made the following significant improvements , making the company X-ray fluorescence analyzer to maintain a leading position .
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
+86-18620722006
instruments@instrumentstrade.com