1
Specifications | Film Test |
Test Range | 0.01 ~ 50,000 cm3/m2·24h·0.1MPa |
Resolution | 0.001 cm3/ m2·24h·0.1MPa |
Temperature Range | 10°C~ 50°C (room temperature 23°C) |
Resolution | 0.01°C |
Temperature Fluctuation | ±0.05°C |
Temperature Accuracy | ±0.3°C (calibration port) |
Vacuum Resolution | 0.01Pa |
Vacuum Accuracy | Displayed Value ±0.2°C (1%~100% of sensor range) |
Vacuum Degree of Test Chamber | < 10 Pa |
Number of Specimens | 3 with independent test results Customization available for other number of specimens |
Specimen Size | Φ97 mm |
Test Area | 38.48 cm2 |
Test Gas | O2, N2, and CO2 (outside of supply scope) Humidifying of Test Gas (customization available) |
Test Pressure | 10KPa ~ 210KPa |
Gas Supply Pressure | 0.5 Mpa ~ 0.6 Mpa (73psi ~ 87psi ) |
Port Size | Φ6 mm PU Tubing |
Instrument Dimension | 710 mm (L) × 350 mm (W) × 630 mm (H) |
Power Supply | 220VAC±10% 50Hz / 120VAC±10% 60Hz |
Net Weight | 100 kg |