The osmotic pressure dew point meter is a dew point method is used to determine the sample osmotic pressure value of the instrument.Compared with the traditional instrument, sample no special processing, can direct determination of high viscosity, containing suspended particles or sample a variety of solvents.In addition to the ordinary liquid samples, osmotic pressure dew point meter can also direct measurement of plant and animal tissue slice, solid samples, application field is very wide.
Features:
Accurate in clinical scope, error less than 1% after 30 years of clinical trials to prove: Vapor osmotic pressure meter is far superior to other similar measuring equipment.
Menu driven streamlined user interface can quickly enter the menu options and commonly used functions
User's choice of measurement model: a single sample measurement mode, automatic repeat mode can measure multiple times on the same sample, the average model can get more accurate data, began to delay model for those who need a longer balance time of sample design.Optional four languages: English, French, German and Spanish
Economic durable just a Vapor osmotic pressure gauge, do not have to buy any accessories or reagent can complete measurement, durable.Throw in 3 q on the manual file, the user can operate at any time certification, no additional cost.
Easy calibration just press a button can automatically set the calibration parameters.
Self-checking built-in self-checking function, on the inspection.
Data storage and output can store up to 16 records, RS - 232 and USB interface can be connected directly to the computer and print, convenient data output.
Automatic cleaning equipped to ion water bottles and waste liquid bottle, automatic cleaning thermocouple head, can reduce the maintenance steps, improve equipment performance.
Technical parameters:
10 microliter sample weight: (standard)
2 ul samples need to use the AC - 063 sample pool, 20 ul samples need to use the AC - 064 sample pool, about 60 ul sample need to use AC - 065 sample pool
Measuring range: generally at room temperature is 20 ~ 3200 mmolkg (5600).
Room temperature: 20 ~ 3600 mmolkg xr (5600)
Measuring time: 90 seconds
Resolution: 1 mmolkg
Repeatability: 20-1000 mmolkg mmolkg 2 or less
1000-3200 mmolkg: 0.5% or less (read)
Linear: the environment temperature in the 20 to 25 degrees, the entire range, reading value plus or minus 5%
In the calibration range 100-1000 mmolkg for reading: plus or minus 1%
Show: 240 x128 LCD display
Operating temperature: 15 ~ 37 degrees, the maximum relative humidity is 85%.(note: at constant temperature calibration for instruments)
Calibration: use WESCOR Opti - mole TM calibration liquid automatic calibration
Data output: RS - 232, USB interface
Power supply: 100-240 - v, 50 to 60 hz, the largest 40 watts
High volume (X width X length) : 28 cmx28cmx36cm
Weight: 6.8 kg