ZY6046B-80L-ZN Thermal Shock Test Chamber Introduction
The thermal shock test chamber, also known as the thermal shock test machine, utilizes low temperature and high temperature cold storage hot water tanks to switch between hot and cold air with cylinder valves according to the action requirements, thereby achieving rapid temperature impact effect, balanced temperature control system (BTC) + The specially designed air supply circulation system controls the SSR in the PID mode, so that the heating amount of the system is equal to the heat loss amount, so it can be used stably for a long time. Technical Parameters
Impact temperature range: -40 ° C ~ 150 ° C
High temperature bath temperature setting range: +60 ° C ~ +200 ° C
High temperature bath heating time RT (room temperature) ~ +200 °C takes about 35 minutes (room temperature is +10 ~ +30 °C)
Low temperature bath temperature setting range: -60 ° C ~ -10 ° C
The low temperature bath cooling time RT (room temperature) ~-60 °C takes about 90 min (room temperature is +10 ~ +30 °C)
Laboratory temperature: -40 ° C ~ 150 ° C
Temperature fluctuation: ±1.0 °C.
Temperature uniformity: ±2.0 °C.
Impact recovery time: -40 ~ +150 °C about 5min (no load), (high and low temperature impact constant temperature time is more than 30min).
Internal dimensions (approx.): W500×H400×D400mm
External dimensions (approx.): W1600xH1850xD1600mm (subject to actual size)
Introduction
The thermal shock test chamber, also known as the thermal shock test machine, utilizes low temperature and high temperature cold storage hot water tanks to switch between hot and cold air with cylinder valves according to the action requirements, thereby achieving rapid temperature impact effect, balanced temperature control system (BTC) + The specially designed air supply circulation system controls the SSR in the PID mode, so that the heating amount of the system is equal to the heat loss amount, so it can be used stably for a long time.