ZR-3930B type ambient air particle sampler Introduction
ZR-3930B type ambient air particle sampler (A type, 10 film; B type, 20 film) (automatic film change), used for air automatic station PM10, PM2.5 data comparison quality control. It is equipped with a standard 16.67L/min cutter (compatible with US EPA standards). It is mainly used to sample PM10/PM2.5 (fine particles) in ambient air, which can automatically change the membrane and automatically sample unattended all-weather.
Executive standard
HJ93-2013 Environmental air particulate matter (PM10 and PM2.5) sampler technical requirements and test methods
HJ618-2011 Determination of ambient air PM10 and PM2.5
HJ656-2013 Technical specification for manual monitoring of airborne particulate matter (PM2.5) (gravimetric method)
JJG943-2011 Total Suspended Particle Sampler
Q/0212ZRB012-2016 Ambient air particulate sampler Features
It can realize automatic replacement of continuous 20 filter membranes, real-time measurement of filter temperature and humidity, ambient atmospheric pressure, temperature, humidity, and automatic constant temperature preservation of standard samples;
The 5.6-inch color LCD high-definition touch screen is easy to operate;
Optional GPRS communication module, real-time query of sampling status, active fault alarm query, real unattended sampling;
The overall casing is designed to be waterproof and dust-proof;
Lightweight aluminum aluminum folding tripod with good support and quick installation for easy carrying;
Built-in high negative pressure brushless sampling pump, when the flow rate is 16.7L/min, the filter resistance is overcome >35kPa;
The sampling pump controller has air circuit blocking and low flow protection;
The automatic transmission module with independent patent technology can smoothly drive the sample;
High life, maintenance free, can work normally at -40 °C;
Sampling process, full-wind design, the temperature difference between the inside and outside of the machine is less than 5 °C;
Adopt high-precision, corrosion-resistant, high-humidity electronic flowmeters;
Built-in large-capacity data storage, with instantaneous data storage, support USB data export;
With automatic leak detection function;
The whole machine is compact in structure, light in weight and small in size;
Power-off memory function, automatic sampling after incoming calls;
AC and DC power supply for power supply; (optional)
Wind direction wind speed sensor with automatic heating and data logging. (optional) Technical parameter
Flow stability: 1h better than ±2.0%
Flow range: 16.7L/min
Running time under high flow and resistance: 2-8h
Pumping pressure: 16.7L/min flow rate, overcome filter resistance greater than 35kPa
Service life under maximum flow and resistance: ≥5000h
Maximum flow error: ±2.0%
Introduction
ZR-3930B type ambient air particle sampler (A type, 10 film; B type, 20 film) (automatic film change), used for air automatic station PM10, PM2.5 data comparison quality control. It is equipped with a standard 16.67L/min cutter (compatible with US EPA standards). It is mainly used to sample PM10/PM2.5 (fine particles) in ambient air, which can automatically change the membrane and automatically sample unattended all-weather.
Executive standard
HJ93-2013 Environmental air particulate matter (PM10 and PM2.5) sampler technical requirements and test methods
HJ618-2011 Determination of ambient air PM10 and PM2.5
HJ656-2013 Technical specification for manual monitoring of airborne particulate matter (PM2.5) (gravimetric method)
JJG943-2011 Total Suspended Particle Sampler
Q/0212ZRB012-2016 Ambient air particulate sampler