Introduction:
ZFSC-I industrial online corrosion rate dynamic monitoring device is designed to solve the difficult problem in the technical supervision of the modern high-parameter, large-capacity generator set in the production process of the on-line analysis of the corrosion rate of the furnace tube, is to carry out thermal equipment Dynamic technology diagnoses and prevents the technical guarantee of leakage and explosion of thermal equipment (furnace tubes).
This product analyzes and measures the water vapor medium, supervises and probes the corrosion information of the heating surface material of the boiler in the production process, reveals the whole process of corrosion and timely understands and grasps the effect of corrosion control, and can timely and quickly diagnose the heat. Corrosion of the equipment in the production process, the actual situation and the resulting security risks. Features:
1. IPC operation, industrial touch screen operation;
2. Dual channel sampling, alternating measurement;
3. Automatically protect the waterway system when the water sample is over temperature;
4. The calibration method is flexible and changeable;
5. The dissolved hydrogen and corrosion rate measurements are automatically converted to 25°C. Technical indicators:
Host: IPC
Display: 1280 x 768 touch screen, Chinese display
Measurement range: (0~10), (0~100)μg/L
Resolution: 0.01μg/L
Indicating error: ±3%FS or ±0.5μg/L
Response time (T90) : 5s
Temperature Sensor: Thermistor
Temperature measurement range: (0.0~50.0)°C
Temperature measurement accuracy: ±0.5°C
Temperature measurement resolution: 0.1°C
Temperature compensation range: (5.0~45.0)°C (manual or automatic)
Atmospheric pressure: Measuring range: 0~115kPa
Measurement accuracy: ±1.5FS
Sample Requirements: Temperature Range: (5.0~45.0)°C
Flow range: (50~350)mL/min
Water sample allowed solids: ≤ 5μm (no gel, no grease)
Alarm output: 4 alarm outputs, programmable alarm content
Digital signal output: RS485, RS232 all the way, programmable
Current output: 4~20mA, 4 channels, programmable
Voltage output: 0~10V, 4 channels, programmable
Ambient temperature: (5.0~45.0)°C
Ambient humidity: ≤90%RH (non-condensing)
Storage and transportation temperature: (-20~55)°C (does not include electrode, electrode should be higher than 0°C)
Power Supply: AC(100~240)V,(50~60)Hz
Power: ≤500W
Insulation strength: 1500V to ground, no arcing in 1 minute
Introduction:
ZFSC-I industrial online corrosion rate dynamic monitoring device is designed to solve the difficult problem in the technical supervision of the modern high-parameter, large-capacity generator set in the production process of the on-line analysis of the corrosion rate of the furnace tube, is to carry out thermal equipment Dynamic technology diagnoses and prevents the technical guarantee of leakage and explosion of thermal equipment (furnace tubes).
This product analyzes and measures the water vapor medium, supervises and probes the corrosion information of the heating surface material of the boiler in the production process, reveals the whole process of corrosion and timely understands and grasps the effect of corrosion control, and can timely and quickly diagnose the heat. Corrosion of the equipment in the production process, the actual situation and the resulting security risks.