Model No. | YTH-080 | YTH-150 | YTH-225 | YTH-408 | YTH-800 | YTH-1000 |
Inner chamber size(cm) | 40×50×40 | 50×60×50 | 60×75×50 | 80×85×60 | 100×100×80 | 100×100×100 |
Exterior chamber size (cm) | 60x172x120 | 70x174x130 | 70x172x181 | 100x182x197 | 120x197x197 | 120x197x217 |
Temperature range | -70℃,-60℃, -40℃, -20℃, 0℃~150℃,other temperature can be customized | |||||
Humidity range | 20%RH~98%RH(10%RH~98%RH 5%RH~98%RH)can be customized) | |||||
Temp.humi.fluctuation | ±0.3℃;±0.5%RH | |||||
Temp.humi.uniformity | ±2℃;±3%RH (empty load, 30 minutes after stablization) | |||||
Temp.humi.resolution | 0.01℃;0.1%RH | |||||
Heating&cooling time ℃/min | Heating rate 3℃/min,cooling rate 1℃/min(average rate, non-linear change) | |||||
Exterior chamber material | Cold rolled plate + Powder coating | |||||
Interior chamber material | SUS#304 Stainless steel | |||||
Insulation material | PU | |||||
Circulation system | Centrifugal fan air circulation system | |||||
Heating system | SUS# Stainless steel high-speed heater | |||||
Humidification system | Electrothermal steam humidification | |||||
Dehumidifying | ADP critical dew point cooling / dehumidifying method | |||||
Refrigeration system | Compressor + Finned evaporator + Air(Water) cooled condenser | |||||
Controlling system | Touch screen controller+SSR+PID | |||||
Other components | Vacuum glass observation window Test hole (50mm) Control status indicator Interior light Loading shelf (2pcs, height adjustable) |
|||||
Safety protection device | Overheat protection circuit breaker Humidification system overheat protection Compressor overload protection Control system overload protection Overload indicator Phase sequence protection Water shortage protection High voltage and low voltage protection Over temperature protection Water supply delay protection Fan overcurrent protection |
|||||
Power supply | AC 1Ψ 220V; 3Ψ 380V 60/50Hz | |||||
Customization service | Non-standard or special requirements can be customized. |