1 Sales TS6210 UV fluorescence sulfur analyzer,Manufacturer
Position: Home > > Analytical instruments > Element analyzer > Sulfur nitrogen analyzer > TS6210 UV fluorescence sulfur analyzer

TS6210 UV fluorescence sulfur analyzer

Product ID:FYld019

Price: Please inquire PDF Format
Supply Ability:70 SETS/MON
Port:SHENZHEN
share:
  • Product Introduction
  • Consulting
  • Model NumberTS6210
    Brand NameBEIJINGNUODETAIKE
    Payment TermsT/T, paypal
    TS6210 UV fluorescence sulfur analyzer
    Introduction
    TS6210 Ultraviolet Fluorescent Sulfur Analyzer is an instrument specially designed for samples with low sulfur content. The minimum detection limit is less than 50ppb. Ultraviolet fluorescent sulfur analyzer is particularly suitable for the detection of ultra-low sulfur samples in petrochemical products. 
    Features
    Detection Systems
    The excitation light source is the core component of the UV fluorescence detector. The gas to be detected generates energy level transitions by irradiation with an excitation light source, thereby generating fluorescence. So whether the choice of the excitation light source is directly related to the accuracy of the instrument measurement.
    Most ultraviolet fluorescence sulfur detectors use a DC-driven zinc lamp as the excitation light source. This excitation method is a single-wavelength excitation SO2 detection, the energy signal is discontinuous, and the excitation light energy intensity decays rapidly with time.
    The TS6210 uses a pulsed flashing xenon lamp as the excitation light source. Since the xenon lamp has a continuous spectral output, the fluorescence intensity is stronger, so the detection limit is lower. At the same time, because the xenon light intensity signal is stable, the service life is longer.
    The electrical pulse modulation driving method can quickly start and stop the light source without preheating. The UV lamp can be turned off when not working for a long time, thereby prolonging the service life of the light source. 
    Vertical furnace design
    Catalytic furnace design for easier maintenance 
    Mass flow meter accurately controls gas flow rate
    Fluctuations of the gas flow will have an important impact on the final analysis results, so precise control of the gas flow rate is a prerequisite for accurate separation and suction results. The TS6210 uses a mass flow meter to precisely control the gas flow rate, avoiding the interference of the gas flow rate and pressure fluctuations on the measurement results, and obtaining the highest decomposition efficiency. When the flow rate or pressure exceeds the threshold, the instrument will automatically alarm and shut down to protect itself, providing a reliable guarantee for the safety of the experiment.
    Liquid autosampler
    Adopting volume difference injection technology: can effectively avoid the interference of air bubbles
    High and low sample analysis: different sample volumes can be used for different samples in the same batch of sample analysis
    Two standard curve: Low concentration curve is used for low content samples, making low content samples more accurate. 
    Extended accessories
    High Nitrogen and Low Sulfur Analysis Attachment (HNLS)
    · Chemical reaction method eliminates the effect of NO on sulfur analysis
    Automatic switching of chemical mode
    · Used for analysis of high nitrogen and low sulfur samples such as fuel oil 
    Gas Analysis Attachment (GAS)
    Stepper motor precisely controls the air-tight needle, and the injection volume is continuously adjustable from 0-100ml
    · The entire analysis process is automatically controlled by software to reduce the perceived error
    · A variety of calibration modes can be used to ensure the accuracy of fidelity analysis results 
    LPG analysis accessory (LPG)
    · Quantitative loop accurate quantification
    Mass flow meter accurately controls flow rate
    Software-controlled full-automatic analysis  
    Technical Parameters
    Meet the standard: SH / T0689-2000; GB / T 34100-2017; ASTM D5453
    Required gas: Ar + O2
    Gas flow rate: flow rate range 350-450mgl / min (input pressure ≤0.12Mpa)
    Sampling method: fully automatic liquid sampler
    Furnace temperature: up to 1200 ° C
    Temperature control accuracy: ± 1% FS
    Combustion tube: quartz reaction tube
    Water removal mode: maintenance-free membrane dryer
    Detector: High-sensitivity UV fluorescence detector
    Measuring range: 0.05-10000 (mg / kg)
    Precision: 5ppm≤5%; 50ppm≤2%
    Analysis time: 5min
    Injection volume: up to 100 μL
    Calibration method: external standard, linear type, segmentable calibration

  • TS6210 Ultraviolet Fluorescent Sulfur Analyzer is an instrument specially designed for samples with low sulfur content. The minimum detection limit is less than 50ppb. Ultraviolet fluorescent sulfur analyzer is particularly suitable for the detection of ultra-low sulfur samples in petrochemical products.
     
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
+86-18620722006
instruments@instrumentstrade.com