1
Wavelength Range | 190nm-900nm | |
monochromator | Eliminating aberration C - T monochromator device | |
Spectral Bandwidth | 0.1,0.2,0.4,1.0,2.0nm:Five-speed automatic switching | |
Wavelength Accuracy | ±0.25nm | |
Wavelength Repeatability | 0.15nm | |
Resolution | Better than 0.3nm | |
Baseline drift | 0.005A/30min | |
Performance Specifications | ||
Characteristic concentration (Cu) | 0.03µg/ml/1% | |
detection limit (Cu) | 0.006µg/ml | |
The burner | Titanium burner | |
precision | RSD≤1% | |
spray | Efficient glass atomizer | |
Atomizing chamber | Corrosion-resistant materials spray chamber | |
Position adjustment | Best flame burner automatically set the height and longitudinal position | |
Safety Measures | With a variety of automatic protection function | |
Graphite furnace analysis | ||
Feature amount (Cd) | 0.5×10-12g | |
detection limit (Cd) | 1×10-12g | |
Precision | RSD≤3% | |
Heating temperature range | Room temperature ~ 2650 °C | |
Heating temperature control mode | Dry ashing stage power control atomization stage light control maximum power mode | |
Setting the heating conditions | Maximum nine programs slopes warming. Ladder warming. Maximum power heating up | |