TA-100 Total Organic Carbon Analyzer Introduction
The TA-100 total organic carbon analyzer uses a modular design, which makes assembly and replacement of parts extremely simple and easier to maintain. TA-100 Total Organic Carbon Analyzer is widely used in quality control of drinking water, industrial water, domestic sewage, production wastewater, etc., and monitoring of rivers, lakes, oceans, surface water, etc. Features
1. The signal processing system with independent intellectual property rights has great advantages such as online setting, real-time monitoring, self-calibration, flow rate control, etc., and guarantees superior instrument performance;
2. The advanced weak current design method is used inside the instrument for safe operation;
3. According to the different properties of different samples, set different control temperatures to completely dissolve the samples and make the measurement data more accurate and reliable;
4. The power of the refrigeration module is adjusted according to the amount of sample injection, which improves the dehydration efficiency of the entire system and ensures the dryness of the gas entering the detector;
5. The development and application of real-time monitoring software, detecting the operating status of the instrument at all times, comparing the differences between the measured parameters and the set parameters. Once the critical setting is exceeded, the system automatically alarms and provides a processing method;
6. Apply the automatic leak detection control system to detect the gas flow in real time, prevent misoperation, improve the performance of the instrument in all directions, and ensure the safety of the operator and the instrument to the greatest extent;
7. Application of flow rate control signal processing technology to shield the effects of flow rate fluctuations; ensure stable and accurate readings;
8. The TIC reaction cell integrates heating, cooling, dehydration, and liquid level monitoring into one, which makes the airflow and liquid flow paths simple, saves a lot of analysis time, and widens the detection limit;
9. The TOC detector uses a 24-bitd digital solution to expand the monitoring range. The control system uses 32bin signal processing technology, which greatly improves the performance of the instrument. Technical Parameters
Model: TA-100 (wet method)
Detector: NDIR (non-dispersive infrared detection)
Measurement items: TC, TIC, TOC, NPOC
Digestion principle: UV / persulfate oxidation
Operation method: computer control
Application object: Water sample, gas (optional)
Gas requirements: Nitrogen: ≥99.995%
Measurement range: 0-1000mg / l
Detection limit: 5µg / l
Reproducibility: 3%
Maximum salt tolerance: 85g / l
Power: 200 ± 10V AC, 50 / 60HZ, 1KW
Size: 430 * 455 * 440mm
The TA-100 total organic carbon analyzer uses a modular design, which makes assembly and replacement of parts extremely simple and easier to maintain. TA-100 Total Organic Carbon Analyzer is widely used in quality control of drinking water, industrial water, domestic sewage, production wastewater, etc., and monitoring of rivers, lakes, oceans, surface water, etc.