- Model NumberSYD-261D
- Brand NameSHCJDZYQ
- Payment TermsT/T, paypal
Instrument overview:
This instrument is designed and manufactured in accordance with the requirements specified in the standard GB/T 261-2008 "determination of flash point binsky-martin closed cup method" of the People's Republic of China. This method is applicable to determine the flash point of closed cup method for petroleum products under specified conditions at the lowest temperature when the mixture of steam and air heated to the cup contact flame.
Product features:
1. Color LCD screen display, all-chinese man-machine dialogue interface, no identification keyboard, and prompt menu-based input for parameters such as prevalue temperature, sample number, atmospheric pressure and test date.
2. The simulated tracking shows the function curve of temperature rise and test time, and has the function of Chinese error operation software to prompt modification, match test date, test time and other parameter prompt functions.
3. Equipped with the standard rs-323,485 computer interface, the lower computer can store 100 sets of historical data. Connected with the computer, the data can be stored in large capacity and stored for a long time.
4. Automatically correct the influence of atmospheric pressure on the test and calculate the modified value.
5. Differential detection and automatic correction of system deviation.
6. The opening of the cover, ignition, detection and printing of data are completed automatically, and the test arm rises and falls automatically.
7. The gas flame is ignited by electronic ignition, which can be ignited by pressing the button.
8, humanized shape design, beautiful, safe, easy to operate.
The biggest characteristic of this instrument is: the color display screen displays all kinds of setting parameters and test parameters, the whole test process is completed automatically, and the test results are printed automatically; Equipped with computer interface, can achieve communication with the computer; The gas flame is ignited by electronic ignition and can be ignited automatically by pressing the button.
Technical indicators:
1. Power supply: AC 220V 10%, 50Hz;
2, temperature measuring range: at room temperature to 300 ℃;
Repeatability: 2 ℃ or less;
Reproducibility: 4 ℃ or less;
Temperature display units: 0.1 ℃;
3, sample heating rate: A step (5 ~ 8 ℃ / min;
B steps: (1 ~ 1.5 ℃ / min;
Automatic control, manual adjustment.
4. Stirring speed: (1) (90 ~ 120) RPM, suitable for test step A.
(2) (250 10) RPM, applicable to test step B.
Automatic control, manual fine-tuning.
5. Ignition mode: gas flame (ignited by electronic ignition);
6, test conditions: temperature (10 ~ 40 ℃;
Relative humidity: 80%;
7. Power consumption: no more than 500W
8. Overall size: (when the lifting arm is not lifted) 520mm * 360mm * 310mm (length * width * height)
(when lifting the lifting arm) 520mm * 360mm * 420mm (length * width * height)