1 Sales SDY-261-1 Closed Cup Flash Point Tester,Manufacturer
Position: Home > > Industry specific instrument > Oil special analytical instrument > Flash / ignition apparatus > SDY-261-1 Closed Cup Flash Point Tester

SDY-261-1 Closed Cup Flash Point Tester

Product ID:HZSYsdy007

Price: Please inquire PDF Format
Supply Ability:200SETS/MON
Port:SHENZHEN
share:
  • Product Introduction
  • Consulting
  • Model NumberSDY-261-1
    Brand NameYIMA
    Payment TermsT/T, paypal
    Summary
    This instrument is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point - Pensky-Martens closed cup method and ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester. It is used to make determination of the closed cup flash point of the petroleum products. 
    I. Main technical features
    1.Single chip microcomputer control technique.LCD screen displays.English man-machine interaction interface.The heating power is continuously step-less adjustable.The rate of heating up is exact.Easy setting for parameters.
    2.The LCD screen has prompting menu,prompting input operation interface.It shows the setted parameters,real-time temperatures and other test parameters.Press the “record”button when the flash point appears.The screen will show and reserve the flash point.
    3. The structure is newly design and small.Stainless steel table-board looks clean and beautiful.It can be used as long as equipped with gas or other civilian fuels.It is easy to use and test results exact. 
    II. Main technical specifications
    1.Power supply:AC( 220±10% )V, 50Hz.
    2.Heating device:(1) The furnace is silicon carbide material. Power is 600W.
                             (2) The heating power is adjustable from 0 W to 600W.
    3. Heating rate:Procedure A: (5~6)°C/min
                                   Procedure B :(1~1.6)°C/min
                          Automatically controlled
    4. Stirring device:(1)Stirring motor: BYGH101 stepping motor
                                        (2) Driving mode: flexible shaft
                              (3) Shaft size: 8mm×40mm             
    5. Stirring rate:(1) Procedure A: (90~120)RPM
                                   (2) Procedure B: (250±10)RPM
    6. Igniting device: (1) Igniting source: gas(or other civilian fuels,the same below)
                               (2) Electric ignition. Gas flame diameter:3.2mm~4.8mm
    7. Temperature testing probe:PT100
    8. Ambient temperature:≤ 35°C
    9. Relative humidity:≤ 85%

     
     
  • Summary
    This instrument is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point - Pensky-Martens closed cup method and ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester. It is used to make determination of the closed cup flash point of the petroleum products. 
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
+86-18620722006
instruments@instrumentstrade.com