RPB-785-1.5T-SS 785nm Raman probe Introduction
The RPB-785-1.5T-SS Raman probe is used in conjunction with a 785nm semiconductor laser and fiber spectrometer to achieve Raman spectroscopy measurement applications. Equipped with different sampling brackets, it can measure solid, liquid or powder samples. OD8's optical interference filter effectively guarantees the filtering of Rayleigh signals. Features
1.Raman spectrum range: 150cm-1-4000cm-1
2.Effectively filter Rayleigh signal, cut-off depth> OD8
3.Start beam consistency: Start <150cm-1 Technical parameters
Dimensions: 102 × 30 × 13 mm
Exit fiber: 105μm core diameter, VIS-NIR band
Receive fiber: 200μm core diameter, VIS-NIR band
Armor specifications of the combined part: Φ8.5mm, PVC black outer armor, metal double buckle tube smoothed inside
Optical fiber (combined section): 115cm ± 2cm
Optical fiber (split beam section): 30cm ± 2cm
Wire harness: N.A.
Trigger device: N.A
Light path control: N.A.
Connect the laser terminal: SMA905
Connect spectrometer terminal: SMA905
Applicable laser wavelength: 785nm
8.5mm protective tube can withstand tensile force: 50N
785 Rayleigh scattering cut-off depth: OD8
Raman spectrum working range: 150cm-1-4000cm-1
Probe working distance: 7.5 mm (standard)
Probe overall exit coupling efficiency: ≥80%
Working temperature: 0 ~ 80 ℃
Working humidity: 5% ~ 80%
Introduction
The RPB-785-1.5T-SS Raman probe is used in conjunction with a 785nm semiconductor laser and fiber spectrometer to achieve Raman spectroscopy measurement applications. Equipped with different sampling brackets, it can measure solid, liquid or powder samples. OD8's optical interference filter effectively guarantees the filtering of Rayleigh signals.