M15 Automatic ELISA analyzer Introduction
The instrument used to read the final ELISA absorbance values. Sampling, testing, values, calculation, printing as one of the automatic ELISA analytical instruments.It can be used to detect all ELISA method. Features
1)No specific restrictions on agents
2)Automatic measurement, automatic calculation, automatic display calculation
3)Large-screen user interface, the user can edit the input detection unit and inspection personnel and other information
4)Absorbance detection, qualitative detection and quantitative detection, quantification There are several calculation methods:
Single-point method, point method, linear regression, logarithmic regression, exponential regression, power regression, regression ratio, the ratio of semi-logarithmic regression, LOGIT, quadratic curve, the ratio of semi-logarithmic law
5)arbitrarily set 96 visual layout, blank bits, control bits, sample location, standard grade
6)Eight-channel detection quickly and accurately
7)Automatically print results
8)Reserved port can load networking.
9)Containing sample database, personnel database and log database. Technical Parameters
1.Resolution:0.001Abs(Display),0.0001 Abs(Internal calculations)
2.Stability:±0.005A
3.repetitiveness:CV≤1%
4.Sensitivity:≥0.010A
5.Indication error:±0.015A
6.Channel difference:≤0.020A
7.Suitability:≤0.005A
8.Wavelength Accuracy:±2nm
9.Light source:Tungsten halogen lamps (12V20W)
10.Measuring channel:8 measuring channels simultaneously detected
11.Insulation resistance:Greater than 20MΩ
12.Sample form:96-well microtiter plate or other types, bar
13.Wavelength: There are four filters: 450nm, 492nm, 630nm, 410nm; can be increased to 10 wavelengths.
14.Vibration function: You can set whether vibration
15.Interface: RS-232 bi-directional communication port, RJ-45 Ethernet ports, USB interface, SD card interface
16.Display: color LCD
17.Operation: Touch screen operation, mouse operation
18.Printing: Built-in thermal printer
19.Working environment: 10 ℃ ~ 35 ℃; relative humidity of 15% - 85%
20.Storage environment: -20 ℃ ~ 50 ℃; relative humidity ≤93%
21.Dimensions:480mm×350mm×210mm
22.Weight:11kg
23. Power supply:AC 220V±22V,50Hz±1Hz
Introduction