- Model NumberLE100
- Brand NameSZSJYQZKSB
- Payment TermsT/T, paypal
Brief introduction:
The LE100 intelligent particle detector is mainly used for the detection of insoluble particles such as injection, infusion, sterile powder, etc. in the Chinese pharmacopoeia (2015), as well as the detection of particle pollution in infusion apparatus, anesthesia apparatus, medicine packaging materials and other medical instruments. It can also be used for the detection of particles in organic, transparent, oil-based and other liquid samples. The instrument met the standards of Chinese pharmacopoeia (2015) and USP <788> (USA pharmacopoeia). Built-in a variety of special testing standards and user - defined standards to meet the requirements of different industries on liquid particle pollution detection. The instrument provides more specialized and intelligent service functions with stronger compatibility and better operability. The unique LE Assistant software complies with the FDA's 21 cfr-11 "electronic record and signature" rule. Provide professional paperless printed report output, with intelligent automatic calibration function.
Instrument features:
Built-in 2015 edition of Chinese pharmacopoeia, GB 8368-2005, anesthesia instruments and other special testing standards
Precision syringe sampling device, sampling volume **, stable speed l built-in calibration curve, in line with iso-21501 instrument calibration requirements
Intelligent automatic detection and calibration functions, instrument performance judgment and calibration
Using magnetic stirring method, stirring speed can be adjusted
USB interface directly export data, paperless electronic printing report
Powerful application development support.
Convenient connection to external equipment such as laboratory information management system
LE -- Assistant software realizes PC control and provides professional data processing
Comply with FDA's 21 cfr-11 "electronic record and signature" rules and user classification management
Application:
Chinese pharmacopoeia, USP, EP, JP and other standards of injection drug testing
Cleanliness inspection of medical equipment/parts
Filter efficiency test
Laboratory pure water test
Technical parameters:
Particle size range: 2-400 m
Particle size channel: 24 channels (2 ~ 100 m)
Detection concentration: particles/ml
Accuracy: 10%
Repeatability: RSD<2% (particle number >1000 /ml, sample size 5ml)
Resolution: >90% (2015 Chinese pharmacopoeia), <10% (USP, ISO21501)
Syringe specifications: 1ml, 5ml, 10ml*, 25ml
Sampling speed: 20ml/min
Sampling volume precision: 1%
Minimum sampling volume: 0.2ml**
Sensor material: 316 stainless steel, quartz glass, PTFE, NBR
Power supply: AC 100~240V, 50/60hz
Stirring mode: magnetic stirring (rotational speed adjustable)
Data output: RS232 interface;
USB;
Thermal printer
Display output: 7 "LCD color touch screen
Size: 320(L) 309(W) 394(H) mm
Weight: 11.0 kg
Using the environment: 10 to 40 ℃ temperature;
Relative humidity 0 ~ 95%, no condensation