HWP08-10S liquid continuous combustion tester Introduction
HWP08-10S liquid continuous combustion tester is used to determine whether the liquid can continue to burn when it is heated and exposed to flames under test conditions. It is suitable for liquids that have no crystals or precipitates at room temperature. It conforms to the United Nations Recommendations and Tests on the Transport of Dangerous Goods. And Standard Manual 32.5.2 Test L.2: Continuous combustion test and G / T21622-2008 "Test method for continuous combustion of flammable liquids of dangerous goods".
During the test, the instrument automatically injects the test liquid into the sample tank, applies a standard flame under specified conditions, and determines whether the sample can continue to burn. It is suitable for the production, transportation, identification, and supervision of chemical liquids. Feature
1) Embedded processor, Windows CE operating system;
2) 8-inch LCD touch screen, which can display test status in real time;
3) Set three working modes: "standard mode", "custom mode" and "cleaning mode" for users to choose arbitrarily;
4) Adopt environmental pressure and temperature compensation technology to precisely control heating temperature and high temperature control accuracy;
5) Automatic intelligent sampling module and maintenance-free automatic cleaning function;
6) High-sensitivity photoelectric detection technology, automatic detection of flame burning phenomenon, ensuring detection accuracy;
7) Automatic ignition, positioning and automatic accurate timing;
8) The instrument has automatic fault detection and sound and light alarm functions;
9) The test process is fully automated without manual intervention, and supports parameter setting, data storage, and viewing reports.
Technical Parameters
1) Working environment: 5 ℃ ~ 30 ℃, <85% RH
2) Standard test temperature: 60.5 ℃ / 75 ℃
3) Temperature control accuracy: ± 0.5 ℃
4) Measuring range of ambient atmospheric pressure: 60 ~ 110kPa
5) Measurement accuracy of ambient atmospheric pressure: ± 0.5kPa
6) Heating timing: 60s, 30s
7) Injection volume: 1 ~ 3mL
8) Injection accuracy: ± 0.05mL
9) Timing range: 0 ~ 999s
10) Timing accuracy: ± 0.01s / min
11) Ignition method: spark plug
12) Sample bottle capacity: 20mL
Introduction
HWP08-10S liquid continuous combustion tester is used to determine whether the liquid can continue to burn when it is heated and exposed to flames under test conditions. It is suitable for liquids that have no crystals or precipitates at room temperature. It conforms to the United Nations Recommendations and Tests on the Transport of Dangerous Goods. And Standard Manual 32.5.2 Test L.2: Continuous combustion test and G / T21622-2008 "Test method for continuous combustion of flammable liquids of dangerous goods".
During the test, the instrument automatically injects the test liquid into the sample tank, applies a standard flame under specified conditions, and determines whether the sample can continue to burn. It is suitable for the production, transportation, identification, and supervision of chemical liquids.