HTF Series Pressurized Thermal Balance Thermogravimetric Analyzer Introduction
The HTF series thermogravimetric analyzer uses a new and unique design, which is a new product that can perform thermogravimetric analysis under high pressure and water vapor conditions. High-precision pressure control and accurate water vapor flow control provide users with better services in the fields of high-pressure analysis of component changes in coal, petroleum, and biomass, process optimization, and quality control. Features
Integrated design of machinery, circuit and atmosphere control.
The furnace body is automatically lifted and positioned accurately, and the measurement repeatability is high.
Large-screen LCD display, displaying instrument status and data all the time.
Water vapor generator with high-precision flow control.
The mass flow meter precisely controls the gas flow.
Data measurement and drawing are done automatically. Powerful software data processing functions (TG extrapolated temperature, 1%, 5%, 50% point temperature, TG extrapolated termination temperature, oxidation induction period, substance purity, data comparison, establishment of standard database).
Software calibration for temperature, quality and other indicators. Technical Parameters
Temperature range: room temperature -1000 ℃
Balance sensitivity: 1 / 10,000
Temperature accuracy: ± 0.1 ℃
Pressure range 0.1 ~ 5Mpa
Adjustable heating rate: 0.1 ℃ / min-80 ℃ / min
Pressure control accuracy 0.02Mpa
Balance measurement range: O-5000mg
Crucible volume: 1 mI-2m1
Introduction
The HTF series thermogravimetric analyzer uses a new and unique design, which is a new product that can perform thermogravimetric analysis under high pressure and water vapor conditions. High-precision pressure control and accurate water vapor flow control provide users with better services in the fields of high-pressure analysis of component changes in coal, petroleum, and biomass, process optimization, and quality control.