Introduction:
FLOM EDI Series FBZ-E Standard Reagent Lab Ultra Purifier is mainly used in atomic absorption (AA), atomic fluorescence (AFS), high performance liquid chromatography (HPLC), mass spectrometry (MC), inductively coupled plasma Buffer preparation, chemical analysis, and verification of emission spectroscopy (ICP); Features:
1. The unique membrane processing technology and accessories in the United States can adapt to different hardness influent water sources; pure water and ultrapure water can be prepared at the same time;
2. Equipped with imported EDI module, the water quality of the produced water is better and more stable, and it can effectively extend the service life of consumables;
3. The advanced source water clean pretreatment Wtr Res-Clean system can prolong the service life of consumables and ensure the normal operation of the instrument for a long time.
4. Reverse osmosis system automatic, manual, regular flushing function, effective protection to extend the RO membrane life;
5. Chinese and English language selection function to facilitate your language usage habits;
6.Multi-function FLOM control system has functions of waterless, consumables failure, expiry of consumables, water leakage, full water alarm, automatic fault detection function, over-standard alarm function for water quality parameter setting, and storage function with various operating data, and Customer password protection function to ensure the operation of data security;
7. The unique three-pump design and fully-closed pipe design automatically circulate during non-water intake to maintain water quality.
8. 5.1-inch LCD large screen can display all kinds of operating parameters intuitively, including inlet water, RO water, UP water three-way water quality and temperature and other parameters;
9. Automatic continuous water production, real-time running status display, adjustable operating parameters;
10. The water system and the circuit system are completely separated, completely eliminating safety hazards;
11. Remote information alarm and water leakage cut off the water source rem-larm module to ensure the safety of the instrument when you are away from the instrument;
12. Wireless control module, remote access to RO / UP water;
13. RS-232 communication module, can connect the printer, print water quality parameters; Technical Parameters:
Product model: FBZ(40-120)01/02
Working conditions: Inlet water: municipal tap water; Source water pressure: 0.15 Mpa-0.5Mpa; Inlet water temperature: 5-40°C
Conductivity (μS/cm) ≤: 0.067
Resistivity (MΩNaN) ≥: 15
TOC content (ppb): TOC≤10
Particulate Matter (0.22μm) : <1/ml
Bacteria (CFU/ml) : <1
Water withdrawal rate (L/min) : 1-2
Water production flow rate (L/h): 40-120
Metal cation (ppb): Fe, Al, Cr, Ni, Cu<0.005ppb, Zn, K<0.02ppb, Na<0.01ppb
Anions (ppb): Cl-, SO42-<0.01ppb, NO2-, NO3-<0.02ppb
Soluble Silicon SiO2/(mg/L) : <0.01
Absorbance (254nm2.5px pathlength): <0.001
Dimensions (mm) : 430*500*560 (length * width * height)
Weight: 40kg
Introduction:
FLOM EDI Series FBZ-E Standard Reagent Lab Ultra Purifier is mainly used in atomic absorption (AA), atomic fluorescence (AFS), high performance liquid chromatography (HPLC), mass spectrometry (MC), inductively coupled plasma Buffer preparation, chemical analysis, and verification of emission spectroscopy (ICP);