1 Sales EFZ-203C online residual chlorine monitor,Manufacturer

EFZ-203C online residual chlorine monitor

Product ID:HSWJyl012

Price: Please inquire PDF Format
Supply Ability:50SETS/MON
Port:SHENZHEN
share:
  • Product Introduction
  • Consulting
  • Model NumberEFZ-203C
    Brand NameHEFEIENFAN
    Payment TermsT/T, paypal
    EFZ-203C online residual chlorine monitor
    Introduction
    The online residual chlorine monitor uses imported components and a permeable membrane head based on the latest polarographic analysis technology and advanced production technology and surface mount technology. Using this series of advanced analytical techniques to ensure the long-term stability and reliability of the instrument. It has Chinese menu operation, 485 communication and other functions. It is used for continuous monitoring and control of residual chlorine content in aqueous solution for drinking water treatment plants, drinking water distribution networks, swimming pools, cooling circulating water, and water treatment projects. 
    Features
    1.Large screen dot matrix LCD display, Chinese menu operation.
    2.Multiple parameters are displayed at the same time: residual chlorine value, hypochlorous acid value, PH value, temperature, output current, etc. are displayed at the same time, which is intuitive and easy to read, and has a range over limit prompt.
    3.There is a factory reset function and a slope correction function.
    4.Manual/automatic PH compensation and temperature compensation.
    5.Communication function (optional): It has RS-485 communication interface (partially compatible with MODBUS protocol), and the residual chlorine value corresponding to 4~20 mA current output can be set arbitrarily.
    6.Optical isolation 4-20mA current output.
    7.The hysteresis can be set arbitrarily to avoid frequent switching of the switching relay. It has the function of setting the switching off and closing.
    8. Watchdog function: Make sure the meter does not crash.
    The core devices are all from famous foreign brands.
    9.Power failure protection > 10 years. 
    Technical Parameters
    Measuring range: hypochlorous acid (HOCL): 0-10.00 mg/L, residual chlorine: 0-20.00 mg/L, pH: 0-14 pH,
    Temperature: 0~60°C
    Resolution: 0.001 mg/L, 0.01 mg/L, 0.01 pH, 0.1 °C
    Accuracy: better than ±1% or ±0.01 mg/L, ±1% or ±0.01 mg/L, ±0.02 pH, ±0.5°C
    Manual/Auto: PH compensation function (0-14) and temperature compensation function (0~60°C).
    Control interface: Two sets of ON/OFF relay contacts, which are divided into high-point and low-point alarm signals.
    Signal isolated output: Photocoupler isolation protection 4 ~ 20mA signal output
    Relay: Relay hysteresis is arbitrarily set, relay load 10A 220VAC
    Working conditions: Ambient temperature is 0~60°C, relative humidity ≤90%
    Output load: Load <500Ω (0-10mA), load <750Ω (4-20mA)
    Input impedance: ≥1×1012Ω
    Working voltage: 220VAC±10%, 50/60Hz
    Size: 96 × 96 × 115 mm
    Opening size: 91×91mm
    Weight: 0.9Kg

  • Introduction
    The online residual chlorine monitor uses imported components and a permeable membrane head based on the latest polarographic analysis technology and advanced production technology and surface mount technology. Using this series of advanced analytical techniques to ensure the long-term stability and reliability of the instrument. It has Chinese menu operation, 485 communication and other functions. It is used for continuous monitoring and control of residual chlorine content in aqueous solution for drinking water treatment plants, drinking water distribution networks, swimming pools, cooling circulating water, and water treatment projects. 
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
+86-18620722006
instruments@instrumentstrade.com