Position: Home > > Industry specific instrument > Textile industry special equipment > Other types of textile equipment > DRK-1000T Mask filter material performance test bench for Mask performance test

DRK-1000T Mask filter material performance test bench for Mask performance test

Product ID:HZFZqt0032

Price: Please inquire PDF Format
Supply Ability:50 SETS/MON
Port:SHENZHEN
share:
  • Product Introduction
  • Consulting
  • Model NumberDRK-1000T
    Brand NameDERUIKE
    Payment TermsT/T,paypal
    DRK-1000T Mask filter material performance test bench
     
    Introduction:
    The DRK-1000T mask filter material performance test bench is used to quickly and accurately detect the efficiency, resistance, filtration rate and flow of various flat materials, such as glass fiber, PTFE, PET, PP melt-blown composite air particulate filter materials, etc. Performance.
    Standards compliant:
            GB 2626-2019 Respiratory protection Self-priming filter anti-particulate respirator
             GB 19082-2009 Technical requirements for medical disposable protective clothing
             GB 19083-2010 Technical requirements for medical protective masks
             GB/T 32610-2016 Technical specifications for daily protective masks
             YY 0469-2011 Medical surgical mask
             YY/T 0969-2013 disposable medical mask
             EN1822-3:2009 High efficiency air filter (sub-high efficiency, high efficiency, ultra high efficiency) Part 3: Filter paper test
             ISO 29463-3:2011 High efficiency air filter and filter material Part 3: Filter paper test
             IEST-RP-CC021.3:2009 HEPA and ULPA filter material test
             JG/T22-1999 General ventilation air filter performance test method
             ANSI/ASHRAE 52.2-2012 General ventilation air filter diameter gauge efficiency test method
             EN779-2012 (air filters for general ventilation-determination of filtration performance)
             JISB9908-2011 (Test methods for air filter devices and electric air cleaners for ventilation).
    Features:
    1. The pressure difference measurement adopts imported brand high-precision pressure difference transmitter to accurately measure the resistance of the sample.
    2. The efficiency test uses two well-known brand laser particle counters to detect the upstream and downstream particle concentration of the sample at the same time to ensure the authenticity and accuracy of the upstream and downstream use of the sample
    3. The fogging system adopts Laskin nozzles, and the release of polydisperse particle size (monodisperse particle size is optional), and the fogging concentration adjustment is fast and stable
    4. Touch screen control, simple and intuitive operation
    5. The test result is automatically calculated and displayed by the system and printed out
    6. Data port: external memory card, data can be exported without worrying about data loss
    Technical Parameters:
    Test flow range: 5100L/min (standard condition 32L/min), ±1%
    Test sample size: 100cm 2, various test fixtures can be customized according to customer requirements.
    Resistance test: measuring range 01500Pa, accuracy up to ±0.025, automatic return to "0" function
    Efficiency test: The efficiency range is 099.999%, and the penetration rate is 0.001%.
    Test particle size: 0.1, 0.15, 0.2, 0.25, 0.3, 0.5, 1.0, 2.0, 5.0, 10.0μ m (select the corresponding sensor according to customer requirements)
    Fog dust source: salt aerosol (NaCL, KCL,), oil aerosol (DEHS, DOP, PAO) and PSL (select when ordering)
    Test time: resistance is tested separately for 10s, efficiency and resistance are tested at the same time for 60s
    Temperature: 050C°, ±0.5C°. Humidity: 0100%RH, ±3%.
    Atmospheric pressure: 8001100hpa, ±0.2%
    Power requirements: AC 220V 50HZ 1.5KW
    Gas source requirements: 0.8MPa, 200L/min
    Dimensions: 700*730*1480mm (length*width*height)
    Product weight: 180Kg
     
  • Introduction:
    The DRK-1000T mask filter material performance test bench is used to quickly and accurately detect the efficiency, resistance, filtration rate and flow of various flat materials, such as glass fiber, PTFE, PET, PP melt-blown composite air particulate filter materials, etc. Performance.
    Standards compliant:
            GB 2626-2019 Respiratory protection Self-priming filter anti-particulate respirator
             GB 19082-2009 Technical requirements for medical disposable protective clothing
             GB 19083-2010 Technical requirements for medical protective masks
             GB/T 32610-2016 Technical specifications for daily protective masks
             YY 0469-2011 Medical surgical mask
             YY/T 0969-2013 disposable medical mask
             EN1822-3:2009 High efficiency air filter (sub-high efficiency, high efficiency, ultra high efficiency) Part 3: Filter paper test
             ISO 29463-3:2011 High efficiency air filter and filter material Part 3: Filter paper test
             IEST-RP-CC021.3:2009 HEPA and ULPA filter material test
             JG/T22-1999 General ventilation air filter performance test method
             ANSI/ASHRAE 52.2-2012 General ventilation air filter diameter gauge efficiency test method
             EN779-2012 (air filters for general ventilation-determination of filtration performance)
             JISB9908-2011 (Test methods for air filter devices and electric air cleaners for ventilation).
     
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
+86-18620722006
instruments@instrumentstrade.com