1 Sales DM2300 X-ray fluorescence multi-element analyzer,Manufacturer
Position: Home > > Analytical instruments > X-ray instrument > X fluorescence spectrometer > DM2300 X-ray fluorescence multi-element analyzer

DM2300 X-ray fluorescence multi-element analyzer

Product ID:FXyg037

Price: Please inquire PDF Format
Supply Ability:90 SETS/MON
Port:SHENZHEN
share:
  • Product Introduction
  • Consulting
  • Model NumberDM2300
    Brand NameSHANGHAIAISITE
    Payment TermsT/T, paypal
    DM2300 X-ray fluorescence multi-element analyzer
    Introduction
    DM2300 X-ray fluorescence analyzer (or X-ray fluorescence spectrometer) (English name: DM2300 EDXRF Analyzer) is the company's original DM series X-ray fluorescence calcium iron analyzer, sulfur analyzer, aluminum silicon analyzer, multi-element analyzer, etc. Based on the development of an analytical instrument that has reached the international advanced level. The instrument uses Energy Disperse X Ray Fluorescence (EDXRF) analysis technology to analyze all elements from aluminum (atomic number 13) to uranium (atomic number 92), which can be analyzed for most elements. Coating thickness measurements can also be made up to 1 ppm up to 100%. It has the characteristics of fast analysis, high precision, small human error, simple operation, only one-time investment, no pollution, so it is widely used in various industries such as electrical and electronic, building materials, metallurgy, petroleum, chemical, geological, mining.
    Features
    (1) Using EDXRF physical analysis method, the sample does not contact or destroy the sample, and no other auxiliary materials such as chemical reagents are needed.
    (2) Multi-element simultaneous rapid analysis, generally give the ppm content of Cl chlorine, chromium Cr, arsenic As, selenium Se, bromine Br, cadmium Cd, strontium Sb, strontium Ba, mercury Hg, lead Pb and other elements in a few minutes.
    (3) With X-ray tube excitation, the voltage and current of the high voltage of the light pipe can be adjusted, so that the instrument can achieve the highest excitation efficiency and the best counting rate. .
    (4) With 4 color filters and 4 collimators, it can be automatically converted and combined to achieve the best counting rate and the highest peak-to-back ratio, which has extremely high precision and small detection limit. .
    (5) High-resolution PIN semiconductor detectors are used to effectively avoid mutual interference of elements.
    (6) Using a high-resolution CCD camera, the sample platform can be moved to ensure accurate measurement of the target position.
    (7) Display, operation and data processing by computer, which makes the operation simple and fast and accurate.
    (8) The instrument has a high degree of integration, good reliability and convenient maintenance. Strong environmental adaptability and long-term stability.
    (9) Using the internal standard method and the parting technology with independent intellectual property rights, the sample can be accurately analyzed without standard samples.
    (10) The sample chamber is large in size and can measure large items.
    (11) The price function ratio is low, which is suitable for China's national conditions.
    (12) The design has excellent shielding and protection, no three wastes, and the radiation protection is safe and reliable. 
    Technical Parameters
    (1) Analysis of element types: Cl chlorine, chromium Cr, arsenic As, selenium Se, bromine Br, cadmium Cd, strontium Sb, strontium Ba, mercury Hg, lead Pb, and the like.
    (2) Analysis range of harmful elements and halogen content: 5 to 3000 ppm.
    (3) System analysis time: 300s (typical).
    (4) Analysis accuracy: Under ideal conditions, the absolute accuracy of measurement is 2~3ppm;
    In the normal state, the relative error of measurement is 10%.
    (5) Detection limit: ≤ 2 ppm.
    (6) X-ray tube: High voltage: ≤50kV, current: ≤1mA, both are continuously adjustable.
    (7) Collimator: Ф1mm, Ф3mm, Ф5mm, Ф7mm, a total of 4 can be automatically switched.
    (8) System energy resolution: For X-rays of MnK5.9keV, its full width at half maximum: ≤155 eV.
    (9) Long-term stability: After 1 hour of starting, the stability of the whole machine for 8 hours: the relative standard deviation of peak position and peak area are less than 0.1%.
    (10) Sample chamber volume: length × width × height: 480 mm × 350 mm × 125 mm.
    (11) Conditions of use: Ambient humidity: 5 ~ 30 ° C, relative humidity ≤ 80% (25 ° C), power supply: 220 ± 20V, 50Hz.
    (12) Power consumption of the whole machine: ≤150W.
    (13) Size and weight: length × width × height: 503 mm × 412 mm × 478 mm; weight: 45 kg.

  • Introduction
    DM2300 X-ray fluorescence analyzer (or X-ray fluorescence spectrometer) (English name: DM2300 EDXRF Analyzer) is the company's original DM series X-ray fluorescence calcium iron analyzer, sulfur analyzer, aluminum silicon analyzer, multi-element analyzer, etc. Based on the development of an analytical instrument that has reached the international advanced level. The instrument uses Energy Disperse X Ray Fluorescence (EDXRF) analysis technology to analyze all elements from aluminum (atomic number 13) to uranium (atomic number 92), which can be analyzed for most elements. Coating thickness measurements can also be made up to 1 ppm up to 100%. It has the characteristics of fast analysis, high precision, small human error, simple operation, only one-time investment, no pollution, so it is widely used in various industries such as electrical and electronic, building materials, metallurgy, petroleum, chemical, geological, mining.
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
+86-18620722006
instruments@instrumentstrade.com