DJC-1 wire and cable vertical burning test equipment Introduction
The single wire and cable vertical burning tester is mainly for the flammability of a single wire and cable with a conductor diameter greater than 8 mm (cross-sectional area greater than 0.5 mm2) or less than 8 mm (cross-sectional area less than 0.5 mm2). It is suitable for the research, production and quality inspection departments of electrical and electronic products and their components such as lighting equipment, low-voltage electrical appliances, household appliances, machine tools, electrical machinery, power tools, electronic instruments, electrical instruments, electrical connectors and accessories. Features
The single wire and cable vertical burning test uses a Bunsen burner of a specified size and a specific gas source (propane) to ignite the sample in a vertical state at a certain flame height and a certain flame angle. Test igniting, duration of combustion, and length of combustion are used to assess flammability and fire hazard. Technical Parameters
1. Burner: inner diameter Φ12mm ± 0.5mm (in accordance with GB/T18380.11) and inner diameter Φ9.5mm ± 0.5mm (in accordance with GB/T18380.21) one Bunsen burner
2. Test inclination: 45°
3. Flame height: 20mm ± 2mm to 190mm ± 1mm adjustable
4. Flame time: 0-999.9s ± 0.1s adjustable
5. Flame holding time: 0-999.9s ± 0.1s, automatic recording, manual suspension
6. Combustion gas: 95% propane gas (in general, liquefied petroleum gas can be used instead)
7. Flow pressure: with dual flow meter and pressure gauge (gas and air)
8. Temperature test range: 0~1000°C
9. Flame temperature requirement: the time from 100 °C ± 5 °C to 700 °C ± 3 °C within 45 seconds ± 5 seconds
10. The flame should continuously burn the sample, and the flame burning time T is set in the range of 0.1-999.9 (seconds) continuously;
11. Temperature measuring thermocouple: Φ0.5mm imported armored thermocouple (K type)
12. Test process: automatic control of test procedures, no independent ventilation
13. Gas: use high-purity propane gas or petroleum liquefied gas (user-supplied);
14. Tested wire and cable length: 600±25mm;
15. Tested wire and cable outer diameter: conductor diameter greater than 0.8mm or cross-sectional area greater than 0.5mm2
16. The device should be placed in an environment that is not ventilated and has a temperature of (23 ± 10) °C.
17. The transfer switch can be manually or automatically selected according to actual needs;
18. Gas flow range: 0.1-1L/min;
19. Air flow range: 0.1-1m3/h
20. Combustion torch standard power: 1KW;
21. Power supply voltage: 220V/50Hz
22. Box material: steel plate electrostatic spraying / or stainless steel chassis, no solder joint process
23. Studio size: 300x450x1200mm (0.16 cubic)
Introduction
The single wire and cable vertical burning tester is mainly for the flammability of a single wire and cable with a conductor diameter greater than 8 mm (cross-sectional area greater than 0.5 mm2) or less than 8 mm (cross-sectional area less than 0.5 mm2). It is suitable for the research, production and quality inspection departments of electrical and electronic products and their components such as lighting equipment, low-voltage electrical appliances, household appliances, machine tools, electrical machinery, power tools, electronic instruments, electrical instruments, electrical connectors and accessories.