    Introduction:
    The CLJ-B330 high flow laser dust particle counter is a new type of dust particle counter suitable for measuring the size and number of dust particles per unit volume of air in clean environments. It can quickly verify the cleanliness level of clean rooms and print relevant test results, complying with ISO14644 and the 2010 version of GMP standards. This product adopts a high-resolution large-sized color touch screen, which is easy to operate and simple to use. In addition, the CLJ-B330 high flow laser dust particle counter can synchronously monitor the number and changes of particles in 8 particle size channels, with a large capacity data storage of up to 500000 pieces of data. Typical applications include particle testing, air particle research, indoor air quality assessment in clean rooms, as well as filter performance testing and cleanliness evaluation.
    characteristic:
    1. Testing scope: Level 100 to 300000 (GMP A BCD               
    2. Large capacity data storage for up to 500000 pieces of data
    3. Built in printer that can print counting results
    4. Stainless steel casing, 7-inch high-resolution color touch screen
    5. Allow high sampling concentration: 35000 particles/L
    6. Various purification levels can be set to exceed the standard alarm
    7. Can display eight particle size channels simultaneously; Low battery display (with optional rechargeable battery)
    8. The test result output complies with ISO14644 and the 2010 version of GMP
    9. The counting results can be converted into m3 and real values
    Technical parameters:
    Sampling volume: 28.3L/min (1cfm)
    Eight particle size channels: 0.3, 0.5, 0.7, 1.0, 2.0, 3.0, 5.0, 10.0 (μ m)
    Detection cycle: 1 second to 99 minutes and 59 seconds, optional
    Self purification time: ≤ 10 minutes
    Working environment: Temperature: 10-35 Humidity: ≤ 75%
    Light source and lifespan: Semiconductor laser light source, with a lifespan of over 30000 hours
    Vacuum source: built-in air pump, flow control
    Communication interface: USB.485. Ethernet communication
    Power consumption: 145W
    Power supply: 220V ± 10%, 50Hz ± 2Hz/built-in lithium battery (optional)
    Dimensions (W × L × H): 220 × 285 × 260 (mm)
    Weight: 8.5Kg

  • Introduction:
    The CLJ-B330 high flow laser dust particle counter is a new type of dust particle counter suitable for measuring the size and number of dust particles per unit volume of air in clean environments. It can quickly verify the cleanliness level of clean rooms and print relevant test results, complying with ISO14644 and the 2010 version of GMP standards. This product adopts a high-resolution large-sized color touch screen, which is easy to operate and simple to use. In addition, the CLJ-B330 high flow laser dust particle counter can synchronously monitor the number and changes of particles in 8 particle size channels, with a large capacity data storage of up to 500000 pieces of data. Typical applications include particle testing, air particle research, indoor air quality assessment in clean rooms, as well as filter performance testing and cleanliness evaluation.
Products
Analytical instruments
Physical property test
Environmental monitoring and analysis
Lab general equipment
About us
About this website
Our service
Product purchase
Contact Us

Newsletter

Make Sure you dont miss interesting happenings by joining our newsletter program.

Contact Us
+86-18620722006
instruments@instrumentstrade.com