BJ-2 Double Cup Disintegration Time Limit Instrument Introduction
The BJ-2 double-cup disintegration time limit instrument is a drug detection instrument for testing the disintegration time limit of tablets, capsules, and pills according to the Chinese Pharmacopoeia. Features
1.Double basket working position system measures the disintegration time limit.
2. 2 sets of hanging baskets can be independently disintegrated.
3.Electronic temperature sensor can display and monitor the temperature in various points in the water bath and in the beaker. The use of high-precision digital electronic sensors ensures high accuracy and high accuracy of water bath temperature without calibration.
4.The instrument automatically controls the temperature of the water bath to 37.0 ℃ (specified by the Pharmacopoeia). The preset temperature can be reset at any time.
5.The instrument automatically sets the lifting time of the gondola to 15 minutes, which can also be reset arbitrarily.
6.The computer control technology with single chip microcomputer as the core, intelligently control the two performance parameters of water bath temperature and working time.
7. The instrument has an overheating alarm and automatic protection for monitoring the temperature of the water bath. Technical Parameters
Disintegration instrument temperature prefabricated range: room temperature ~ 50 ℃, display resolution is 0.1 ℃.
Temperature control accuracy: ± 0.3 ℃
Timing prefabrication range: 1min ~ 900min
Time control accuracy: ± 0.5min
Working noise: <60db
Number of lifting baskets: 2 sets
Lifting frequency of gondola: (30 ~ 31) times / min
Lifting distance of gondola: (55 ± 1) mm The minimum distance from the screen to the bottom of the cup: 25mm ± 2mm
Screen aperture: standard configuration 2mm, (0.425mm, 0.71mm, 1mm aperture can be selected)
Power source: 220V50Hz
Overall power: 600W
Dimensions: length * width * height 400mm * 320mm * 440mm
The BJ-2 double-cup disintegration time limit instrument is a drug detection instrument for testing the disintegration time limit of tablets, capsules, and pills according to the Chinese Pharmacopoeia. BJ-2 double-cup disintegration time limit instrument adopts the computer control technology of single chip microcomputer as the core, and intelligently controls the two performance parameters of water bath temperature and working time.