National standards implemented | GB/T 3785-2010 (IEC 61672-1: 2013) 2 GB/T 3241-2010 2 / IEC 61260-1: 2014 Class 2 |
Metrological verification regulation for execution | JJG 188-2017 Sound level Meter, JJG 778-2005 noise Statistics Analyzer, JJG 449-2014 < double Frequency path filter and Fractional doubling path filter > |
Frequency range | 20 Hz~12.5 kHz |
Measuring range | 28~133 dB(A);33~133 dB(C);40~133 dB(Z) |
Order linear range | 105 dB(A) |
Frequency weighting | Parallel A, C, Z |
Time weighting | Parallel (at the same time) F, S, I |
Indicator | 2.6 inch color screen display, resolution 240 x 320, display content is rich, backlight delay is normally open, brightness automatic adjustment, other manual adjustment |
Main measurement index | Lxyi, Lxyp, Lxeq, Lxmax, Lxmin, LxN, SD and 24-hour automatic measurement Ld, Ln, Ldn, etc. Note: x is A, C, Z, y is F,S,I, N is 1 ~ 99 users can choose an integer |
A/D sampling frequency | 48k per second |
Automatic 24-hour monitoring | The measurement time can be chosen between 1 min~1hour per hour, and can be measured continuously for more than 24 hours. |
Storage | Single statistical analysis of 32Mb ~ (3300) band distribution map |
Output interface | 1) ac output, output power: 150 mW, can pick up 8 Ω monitor headphones 2) dc output: output the dc signal proportional to the sound pressure level currently displayed, 15 mV/dB 3) RS232 interface: communicate with the computer, and print the measurement results and relevant charts through the relevant micro-printer 4) USB interface: communicate with computer, upgrade firmware, conform to USB1.1 standard, compatible with USB2.0 standard |
Internal calendar clock | Error less than 1 minute / month, GPS timing |
Power supply | 4 AA alkaline batteries can be continuously used for more than 30 hours to meet the requirements of 24-hour continuous monitoring in environmental monitoring |
Shape dimension (mm) | 260×80×30 |
Quality (kg) | 0.35 |
Work environment | Operating temperature range: - 10 ℃ to 50 ℃;Relative humidity: 20% ~ 90%;Atmospheric pressure: 65 kPa ~ 108 kPa |
Calibration | AWA6223F multi-frequency point calibrator is used to meet the requirements of sound level calibrator in HJ 707-2014 environmental sound monitoring technical specification |
Print function | AH40 dot matrix printer, printing data can be saved for a long time |