AOA-100 atmospheric heavy metals online analyzer Introduction
AOA-100 atmospheric heavy metals online analyzer is an instrument of lanscientific based on on-line detection technology and automatic air particulate matter enrichment technology perfect union, in line with the U. S. EPA IO-3.3 standard method, unattended, long time automatic monitoring. Features
1.The main engine adopts the international standard 19 case, which can be installed on the mobile monitoring vehicle, and can also be installed on the standard cabinet of the air automatic station;
2.Instrument has a very high operating reliability and stability, but also with intelligent fault diagnosis and self exclusion function;
3.No stray light path design, built-in multi filter, automatic switching to achieve low background measurement;
4.Built in calibration module, XRF spectral horizontal and vertical coordinate automatic calibration, calibration frequency can be arbitrarily set, also built a specific internal standard metal source, can realize automatic quality control and alarm;
5.By means of the mass flow controller and the temperature and pressure sensor, the flow rate of the sample is controlled accurately and stably, and the automatic calibration of the flow can be realized;
6.With RS232/485, USB, GPRS, ADSL, Ethernet and other data transmission mode, you can achieve the operating parameters, state and data transmission, monitoring and inspection;
7.System software uses advanced hierarchical modular design, with a strong, flexible data settings, calibration, storage, display, query, analysis, transmission and other functions. Data in line with the "pollution source online automatic monitoring (monitoring) system data transmission standards" (212-2005 HJ/T). At the same time, users can quickly and easily customize the expansion of any other custom functions;
8.Professional multi radiation treatment, measuring the full range of the instrument without any radiation leakage. Technical Parameters
Working power supply: 220V AC±10%, 50Hz±1Hz
Measuring ambient temperature: -40~50°C
Measuring ambient humidity: 0%RH~90%RH
Host ambient temperature: 0 ° C ~ 40 ° C
Host environment humidity: 30% RH ~ 60% RH
Host size: 482mmx 420mm x 600mm, (19" cabinet)
Working mode: Continuous automatic monitoring; one year of trouble-free operation, and fault self-recovery function.
Control mode: Computer control, self-diagnosis and setting function, real-time status monitoring function, and can store monitoring data for more than ten years.
Operation interface: Chinese or English operation interface, touch screen operation, display real-time sampling flow, sampling time, measurement status, heavy metal concentration value (standard state or non-standard state optional setting), content curve and other information.
Installation method: Standard cabinet installation or mobile vehicle installation in fixed station.
Monitoring indicators: TSP/PM10/PM2.5 mass concentration
Simultaneously monitor 30 elements such as Pb (lead), Cd (cadmium), Hg (mercury), and As (arsenic) (other elements can be expanded according to the user's subsequent needs)
Applicable standards: Heavy metals: US EPA IO-3.3 Standard method "X-ray fluorescence spectrometry for the detection of heavy metals in environmental particles"
Fine particulate matter: Technical requirements and test methods for continuous monitoring system of ambient air particulate matter (PM10 and PM2.5)
Measuring range: Heavy metal: 0~100μg/m3
Fine particles: 0 ~ 10mg / m3
Detection limit: Heavy metal: ng/m3 order (sampling time 1 hour, flow rate 16.7L/min)
Fine particles: <1μg/m3 (24-hour average)
Precision: Heavy metal: RSD<1.5% (Pb, Cd, As sampling time 4 h).
Fine particles: within ±5μg/m3 (24 hours)
Sampling flow: (4~20) L/min adjustable
Sampling and analysis time: 10 to 300 minutes, adjustable
Introduction