ALC10 aflatoxin tester Introduction
The ALC10 aflatoxin analyzer integrates a liquid chromatography system, a post-column photochemical derivative module, and a fixed optical path fluorescence detection module. The minimum detection concentration, linear range, repeatability and other indicators are in line with national standards and relevant regulations. Features
In the aflatoxin test, the ALC10 achieves the same level of general-purpose liquid chromatograph with a derivative and fluorescence detector, and can meet the testing needs at half the cost.
The ALC10 has a built-in standard application method that uses a one-button operation, gentle overpressure for real-time monitoring and protection, and the whole machine is durable. Easily control the instrument without the need for the operator's expertise.
The chromatographic workstation adopts a database storage mode and has a comprehensive audit trail function and privilege grading settings to fully guarantee workstation data security and compliance.
The life of the light source is more than 20,000 hours, and the liquid leakage, over-temperature and over-pressure can be monitored and protected in real time, and the whole machine is durable. Technical Parameters
ALC10 machine performance parameters
Retention time repeatability: RSD ≤ 0.2%
Peak area repeatability: RSD ≤ 1.0%
ALC10 infusion pump parameters
Withstand voltage: 45MPa
Flow rate range: 0.001mL/min-10.000mL/min
Flow accuracy: ±0.5% (1mL/min)
Flow precision: RSD ≤ 0.1% (1mL/min)
ALC10 column oven parameters
Temperature indication error: ±1 °C
Temperature stability: ±0.1°C
ALC10 detector parameters
Excitation wavelength: 360nm
Emission wavelength: 440nm
Dynamic baseline drift: ≤ 500 counts/h
Short-term baseline noise: ≤ 50 counts
Minimum detection concentration: ≤ 0.03ng/mL (AFT B1, AFT G1)
≤ 0.01ng/mL (AFT B2, AFT G2)
Linear range: > 103
Introduction
The ALC10 aflatoxin analyzer integrates a liquid chromatography system, a post-column photochemical derivative module, and a fixed optical path fluorescence detection module. The minimum detection concentration, linear range, repeatability and other indicators are in line with national standards and relevant regulations.