A1220 Insulating Oil Gas Chromatograph Introduction
The A1220 gas chromatograph is designed and manufactured according to the methods specified in GB / T 17623 and DL / T 703 standards, and is suitable for analyzing oil-filled electrical equipment (including transformers, reactors, current transformers, voltage transformers, charging bushings). Etc.) Analysis of hydrogen, carbon monoxide, methane, carbon dioxide, ethylene, ethane, acetylene and other gas contents dissolved in insulating oil. Features
High sensitivity, high accuracy, high resolution, high analysis speed, simple analysis method.
One chromatographic column has good separation effect.
One injection, less injection volume; short time for full analysis.
The data is processed automatically by the chromatography workstation. Technical Parameters
Start-up stabilization time: <1.5 hours
Display accuracy: 0.1 ℃
Temperature control accuracy: ≤ ± 0.3 ℃
Baseline noise: TCD≤0.1mv FID≤1 × 10-12A / 30min
Baseline drift: TCD≤0.2mv / 30min FID≤1 × 10-11A / 30min
Sensitivity: TCD≥1000mvml / mg (benzene)
Detection limit: FID≤5 × 10-12g / s
Oven temperature: room temperature to 200 ° C
Detector (FID) temperature: room temperature to 300 ° C
Methanation conversion furnace temperature: 380 ℃
Over-temperature protection: Any temperature exceeding the limit will automatically alarm and power off
Detection index: CH4 0.1PPm
CO 1PPm
CO2 1PPm
H2 5PPm
C2H2 0.1PPm
C2H4 0.1PPm
C2H6 0.1PPm
Dimensions: 620mm × 445mm × 485mm
Weight: 31.8kg
Introduction
The A1220 gas chromatograph is designed and manufactured according to the methods specified in GB / T 17623 and DL / T 703 standards, and is suitable for analyzing oil-filled electrical equipment (including transformers, reactors, current transformers, voltage transformers, charging bushings). Etc.) Analysis of hydrogen, carbon monoxide, methane, carbon dioxide, ethylene, ethane, acetylene and other gas contents dissolved in insulating oil.