A1072 Coulomb Trace Water Analyzer Introduction
A1072 Coulometric method for trace water analyzer: GB7600-1987, GB / T606-2003, GB / T3727-2003, GB / 8350-2001, GB / 8351-2004, GB / T 6023-1999, GB / T 6283, GB / T 11133-89, GB / T 11146-1999, A1072 are based on the Coulomb titration principle and automatically controlled by a microcomputer. They have the advantages of high sensitivity, fast electrolysis speed, short equilibration time, accurate blank deduction, and accurate and reliable analysis results. Applications in petroleum, chemical, power, environmental protection, scientific research and other departments. Features
Control system: SCM and computer compound control.
Zero point correction: The accurate software deducts the blank function, which guarantees that the blank can be accurately deducted within the 10-minute sample enrichment time.
Equilibrium point (titration endpoint) setting: Equivalent to the pH value of the product during acid-base titration, different indicators can be selected. This function can improve the adaptability to different sensitivity reagents, ensuring that the instrument is used in various All reagents can be adjusted to the optimal measurement sensitive area for analysis.
Communication function: It can be connected to an external balance, and can communicate with a computer, store a large amount of data in the computer, and control the instrument with a computer for easy operation.
Analysis accuracy setting function: According to the analysis requirements of different measurement samples, users can freely choose the analysis speed and measurement accuracy.
Analysis result storage function: can store 100 measurement data.
Reagent replacement prompt function: It can prompt the user to replace the reagent that is about to expire.
Fault self-diagnostic function: When there is a fault in the electrolysis part or measurement part of the instrument, the fault is automatically prompted, which is convenient for the user to diagnose the fault. Technical Parameters
Method used: Karl Fischer Coulomb method
Titration (electrolysis) control method and speed: automatic control of electrolytic current, maximum electrolytic current 360mA, maximum titration speed 2mg ∕ min, software integration
Low detection concentration: when the mass of a sample is 10g, 0.00003% (m ∕ m) is 0.3ppm
Detection limit: 3μgH2O
Sensitivity: 0.1μgH2O (instrument resolution)
Accuracy: 3 micrograms of water ± 20%, 10 micrograms of water ± 10%, 100 micrograms of water ± 1%, 100 micrograms of water ± 0.5% or more (based on the absolute water content of the sample X, the relative error should not exceed ± 0.5 multiplied by Divide the cubic root of X by X and multiply by 100%)
Measurement range: 3μg ~ 200 mg
Full Chinese menu operation, intuitive and simple; wide line mini thermal printer, clear handwriting, low noise and fast speed
Working power: AC220V ± 10%, 50Hz;
Relative temperature; 5 ° C ~ 40 ° C:
Relative humidity <80%
Dimensions: 675mmX425mmX285mm
Weight: 5kg
Introduction
A1072 Coulometric method for trace water analyzer: GB7600-1987, GB / T606-2003, GB / T3727-2003, GB / 8350-2001, GB / 8351-2004, GB / T 6023-1999, GB / T 6283, GB / T 11133-89, GB / T 11146-1999, A1072 are based on the Coulomb titration principle and automatically controlled by a microcomputer. They have the advantages of high sensitivity, fast electrolysis speed, short equilibration time, accurate blank deduction, and accurate and reliable analysis results. Applications in petroleum, chemical, power, environmental protection, scientific research and other departments.